Clearing Malfunctions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
177
13
P a r . r e l e a s e
C A U T I O N
T h i s m e a s u r e m e n t i s
n o t v a l i d . P l e a s e
c h e c k a n d r e l e a s e
t h e p a r a m e t e r s .
Internal malfunction.
The measured values
are not valid. The
individual settings on
the device are not
effective.
Can only occur
during
manufacturing.
Disappears as soon
as the cause is
eliminated.
Have the EuroFID serviced
by the manufacturer's
after-sales service.
14
M a i n t . s w i . O N
C A U T I O N
S e v e r a l f a c t o r y
s w i t c h e s a r e s e t .
T h e s y s t e m i s i n
t e s t m o d e . T h i s
m e a s u r e m e n t i s
i n v a l i d .
Internal special
functions are activated.
The measurement is
invalid.
15
A v e r a g e e r r .
S Y S T E M E R R O R
T h e a v e r a g e a r c h i v e s
a r e l o s t . A l l
e n t r i e s a r e d e l e t e d .
A n e w a r c h i v e w i l l
b e c r e a t e d .
Internal malfunction:
The saved average
values are lost. (Does
not affect the current
measured values.)
Can only be deleted
by switching the
device off.
To delete the message:
Switch the EuroFID off and
switch it on again.
16
E r r o r f i l . e r r
S Y S T E M E R R O R
T h e e r r o r a r c h i v e s
a r e g o n e . A l l
e n t r i e s a r e d e l e t e d .
A n e w a r c h i v e w i l l
b e c r e a t e d .
Internal malfunction:
The content of the error
archive has been
lost.(Does not affect
the current measured
values.)
17
C l o c k p a r a E R R
S Y S T E M E R R O R
T h e p a r a m e t e r s o f
t h e c l o c k c h i p a r e
l o s t . P l e a s e
i n i t i a l i z e t h e
p a r a m e t e r s .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
Set the span gas setpoint
value.
Check/set the conversion
factors (
).
Then calibrate the device
(
18
E P R O M E r r o r
S Y S T E M E R R O R
T h e s y s t e m c h e c k h a s
d e t e c t e d a R O M
e r r o r . R e s t a r t t h e
s y s t e m !
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
EuroFID Switch off and on
once more.
If unsuccessful:
Have the
EuroFID serviced by the
after-sales service of the
manufacturer.
19
R A M E r r o r
S Y S T E M E R R O R
T h e s y s t e m c h e c k h a s
d e t e c t e d a R A M
e r r o r . R e s t a r t t h e
s y s t e m !
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
Set the span gas setpoint
value.
Check/set the conversion
factors (
).
Then calibrate the device
(
20
F _ E P R E r r o r
S Y S T E M E R R O R
T h e s y s t e m c h e c k h a s
d e t e c t e d a R A M -
P A R I T Y e r r o r .
R e s t a r t t h e s y s t e m !
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
EuroFID Switch off and on
once more.
If unsuccessful:
Have the
EuroFID serviced by the
after-sales service of the
manufacturer.
21
H A R D W E r r o r
S Y S T E M E R R O R
T h e s y s t e m c h e c k h a s
d e t e c t e d a h a r d w a r e
e r r o r .
22
L C D - D i s p - E r r
S E R V I C E R E Q U I R E M E N T
T h e s y s t e m c h e c k h a s
d e t e c t e d a L C D
d i s p l a y e r r o r .
Internal malfunction.
Measurement is,
however, not
interrupted.
Disappears as soon
as the cause is
eliminated.
EuroFID Switch off and on
once more.
If unsuccessful:
Have the
EuroFID serviced by the
after-sales service of the
manufacturer.
23
B a t t e r y l o w
S E R V I C E R E Q U I R E M E N T
C l o c k b a t t e r y b a c k u p
p o w e r t o o l o w . P l e a s e
c h a n g e t h e b a t t e r y .
T y p e B R - 1 2 2 5 ;
1 2 . 5 m m d i a m e t e r
2 . 5 m m h e i g h t
3 . 0 V v o l t a g e
Warning - possible
device failure.
Measurement is not
interrupted.
Disappears as soon
as the cause is
eliminated.
Have the internal batteries
replaced by the
manufacturer's after-sales
service.
24
T i m e i n p u t
S E R V I C E R E Q U I R E M E N T
T h e t i m e i s l o s t a n d
s t a r t s a t 0 0 : 0 0 : 0 0 .
P l e a s e s e t t h e t i m e .
Internal malfunction.
Measurement is not
interrupted.
Disappears as soon
as the cause is
eliminated.
Reset the time of the
internal clock (
N Spontaneous message
Message text in the error archive
Significance
Duration of the
message
Countermeasures
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...