Instructions for Use 

 

DNPH active gas tubes must be stored at low temperatures. 
Therefore, be sure to take measures to keep them cool while 
carrying them. If transporting them, make sure they are kept at 

°

C or below during transport. 

 

z

 Sampling

 (adsorption in the active gas tube) 

(1) Cut open the aluminum bag with scissors or something similar. 

(Markings on the aluminum bag indicate where to cut.) The bag 
contains the DNPH active gas tubes individually packaged in aluminum. 

Keep any bags not immediately used for sampling stored in a location 

at 4 

°

C or below. 

(2) Just before sampling, cut the small aluminum bag at the cut mark 

indicated with an arrow (can be cut by hand). Then open the zip lock 
seal and remove the DNPH active gas tube. 

Note:  The small aluminum bag includes a notation area for recording the measurement 

point or other information. When writing down, use a pencil, which has little effect 
on measurements, and avoid using an ink pen. 

(3)  Remove the polyethylene plug 

and cap from the DNPH active 
gas tube and use Toalon tubing 
(flexible PVC tubing) or other 
means to connect the active gas 
tube to the pump. Then start 
sampling. Perform measurements 
in accordance with laws and 
regulations applicable for given 
measurement objectives. 

Note:  The outer diameter of the suction 

end of the DNPH active gas tube is 
about 6 mm. 

Note:  To measure concentrations in 

outdoor air environments, Sibata 
recommends using a separately sold 
ozone scrubber connected to the end 
of the DNPH active gas tube to avoid 
the influence of ozone. 

(4) After exposure, promptly reinstall the polyethylene plug and cap on the 

DNPH active gas tube, put the assembly back into the small aluminum 
bag, and seal the zip lock closure. 

Then keep it stored in a location at 4 

°

C or below. 

DNPH active

gas tube

Tube

Pump 

Содержание 080150-076

Страница 1: ...ual describes important precautions for preventing accidents as well as procedures for handling the product z To ensure safety read this operation manual thoroughly before use and use the product corr...

Страница 2: ...ysis Procedure z Preparing Standard Solutions z Preparing Calibration Curves z Analyzing Samples z Calculating Concentrations z Calculation Method in Japan s Ministry of Health Labour and Welfare Guid...

Страница 3: ...hat the information in this operation manual is correct If you discover any errors or omissions however please contact your Sibata representative z The copyright of this operation manual belongs to Si...

Страница 4: ...ymbols are as follows Fully understand the following descriptions before reading the subsequent sections Indicates a potentially hazardous situation which if not avoided could result in serious injury...

Страница 5: ...hibited Do not put DNPH active gas tubes in your mouth Not doing so could result in accidents or fires Doing so could cause injuries or other accidents Mandatory Be sure to keep the product in a locat...

Страница 6: ...es by putting them in contact with silica gel impregnated with 2 4 dinitrophenylhydrazine hereinafter DNPH in a polyethylene tube Then after extracting the aldehydes and ketones with acetonitrile they...

Страница 7: ...recording the measurement point or other information When writing down use a pencil which has little effect on measurements and avoid using an ink pen 3 Remove the polyethylene plug and cap from the D...

Страница 8: ...Prepare standard solutions by diluting commercial DNPH derivatives of aldehydes and ketones with acetonitrile Formaldehyde 4 g HCHO mL Place 1 mL of the formaldehyde DNPH standard solution 40 g HCHO m...

Страница 9: ...ge to the tube and slowly inject 5 mL of acetonitrile Add additional acetonitrile to make 5 mL for use as the sample solution Analyze the sample solution in the same manner as described in Preparing C...

Страница 10: ...results are to be compared to indoor air quality guideline values indicated by the Ministry of Health Labour and Welfare Calculation Formula Formula 3 z C Concentration g m3 of aldehydes in the sampl...

Страница 11: ...i pump MP 30HN 080860 104 Mini pump MP 100HN 020060 5A Volumetric flasks super grade 5 mL set of 10 Product Shelf Life The shelf life of the product is indicated on the outer packaging Confirm that th...

Страница 12: ...about this product or if there is any other way in which we can be of assistance contact your Sibata representative or the vendor where you purchased the product Please also visit our website to cont...

Отзывы: