When a difference develops in the refractive index between the sample and the
Reference in the flow cell, the image on the photo sensor moves horizontally as
illustrated in Fig. 4.3 below ( moves from (a) to (b) ). The electric signals emitted
respectively from the photo sensor change in proportion to the extent of movement
of the image.
Thus, signal corresponding to the difference in refractive index can be obtained from
a difference between the two signal outputs.
(a)
No difference in the
refractive index
(b)
Some difference in the
refractive index
Fig. 4.3 Movement of the image on the photo sensor
4
4
4
4-
-
-
-2
2
2
2 Flow
Flow
Flow
Flow L
L
L
Line
ine
ine
ine
As shown in Fig 4.4 below, the flow line is so designed as to enable replacement of
a reference liquid by only pushing a button.
Fig. 4.4 Flow Line