background image

Asahipak ODP-50 (C8P, C4P series) 

 

 

Manual 

 

4.

 

Column installation

 

4.1 Direction of flow 

Install and use the column with mobile phase flowing in the direction indicated on the column 

tag. 

If the column shows very high resistance to flow, it may be partially plugged up. This is probably 

the end of the service life of the column. Some users have been successful in extending the life 

somewhat by back flushing the column. The outlet of the column is connected to the injector so 

the liquid flows into the normal outlet fitting. The other end, normally the column inlet is placed 

into  a  beaker.  The  pump  is  then  turned  on  at  the  minimum  flow  rate  listed  in  Table  1.  The 

column  is  back  flushed  for  about  one  hour  to  over  night,  or  even  longer,  if  the  operating 

pressure  is  declining.  The  column  then  should  be  retested  according  to  the  procedure  on  the 

page  "INSPECTION  DATA"  provided  with  each  column.  If the  column  is  efficient,  then  one  can 

restore the column to operation in the instrument in the normal manner. 

See also Section 5.1 on filtering and guard columns. 

4.2 Connection tubes 

For columns with diameters smaller than 10 mm id, use connection tubes from the injector and 

to the detector that have the shortest practical length and internal diameter no larger than 0.3 

mm, (0.010 inch) internal diameter. The outer diameter must be 1/16 inch, (0.0625 inch). 

For preparative columns, (columns with internal diameters of 10 mm and larger), it is possible to 

use connection tubes with internal diameter up to 0.5 mm, (0.020 inch). The length should be 

kept as short as practical. 

4.3 Pump 

The use of pumps with strong pulsations in the output flow or pressure may shorten column life, 

and should be avoided. Pulsation can often be reduced by using a pulse damper on the pump 

outlet. 

4.4 Preliminary purge 

Before connecting the column to the system, purge the system and lines with liquid to remove 

any air or particles. 

 

 

Содержание Asahipak C4P series

Страница 1: ...dex HPLC Columns Europe Middle East Africa Russia SHOWA DENKO EUROPE GmbH Shodex Business Konrad Zuse Platz 3 81829 Munich Germany E mail support shodex de Phone 49 0 89 93 99 62 37 www shodex de For technical support please use contact details shown below ...

Страница 2: ......

Страница 3: ...pak ODP columns are compatible with the commonly used reversed phase HPLC eluents between pH 2 to pH 13 The particles are available in two sizes The smallest has a diameter of five micron These are identified by the suffix 50 in the first column of Table 1 The small particles give the highest column efficiency and resolution but at the expense of higher operating pressure The Asahipak 90 series de...

Страница 4: ...nstructed from passivated 316 stainless steel The column end fitting is designed to minimize dead volume and band broadening The depth of the fitting for the inlet and outlet tubes may be different than those used on columns from some other manufacturers If you use a tube with a short nose extending from the fixed ferrule there will be a cavity in the flow path that can lower the column efficiency...

Страница 5: ...3 0 C8P 50 4E 4 6 x 250 5 10 11000 0 5 0 8 1 0 4D 4 6 x 150 5 10 7000 0 5 0 8 1 0 10E 10 0 x 250 5 10 8000 1 5 2 0 3 0 C4P 50 4E 4 6 x 250 5 6 9000 0 5 0 8 1 0 4D 4 6 x 150 5 6 6000 0 5 0 8 1 0 10E 10 0 x 250 5 6 7000 1 5 2 0 3 0 ODP 90 20F 20 x 300 9 17 9000 5 10 12 Table 2 Guard columns Grade Column Dimensions IDxL mm Columns guarded IDxL mm ODP 50G 4A 4 6 x 10 ODP 50 4 6x150 4 6x250 ODP 50G 6A ...

Страница 6: ...INSPECTION DATA provided with each column If the column is efficient then one can restore the column to operation in the instrument in the normal manner See also Section 5 1 on filtering and guard columns 4 2 Connection tubes For columns with diameters smaller than 10 mm id use connection tubes from the injector and to the detector that have the shortest practical length and internal diameter no l...

Страница 7: ... purpose Filtering is not always effective in preventing plugging of the column Experience shows that if column plugging is a frequent problem that the plug is often located at the very top of the column Guard columns are short sacrificial columns that are installed ahead of the main column to protect it When a plug occurs the guard column is replaced 5 2 Degassing Numerous practical problems are ...

Страница 8: ...ure low below 15MPa It is important to check the solubility of mixed solvent buffers before running a solvent program This can be done by making a quick scouting run in a test tube Slowly add the strong eluent to a small quantity of the starting eluent If the solution turns cloudy or white this means a precipitate is forming Other buffer combinations should be tried 7 Elution modes All Asahipak co...

Страница 9: ... degassing may be required to prevent formation of bubbles in the detector At low temperature the viscosity of the mobile phase will increase Care should be taken to avoid exceeding the maximum pressure specification of the column This may require decreasing the flow rate Flow rate is proportional to pressure so cutting the flow in half will decrease the pressure on the column by half 10 Column ha...

Страница 10: ... the initial mobile phase composition In some cases it is desirable to prepare the sample in a solution with an elution strength slightly weaker than the mobile phase Upon injection the sample components will tend to focus in a narrow band on top of the column This can be improve detection limits 12 Measurement of NTP The measurement conditions employed in the determination of NTP are described in...

Страница 11: ...nt is granted or implied by our sales of any material Shodex Asahipak products in this document are for laboratory use only They are not intended for medicinal or food use We assume no responsibility if these products are used for medicinal or food purposes or are misused in any way Shodex Asahipak products are not intended for clinical use While clinical applications may be shown these products a...

Отзывы: