background image

 

8

 

In the  case of immersion type thermocouple measuring mode, Data HOLD [D.HOLD] 

 function does not work. 

6. Other functions 

      

• Automatic cold junction temperature compensation

 (Only thermocouple input) 

        Detects the temperature at thermocouple and connecting terminal of this instrument and constantly 

keeps the reference junction status at 0  or 32  

      

• Auto-Power-Off

 

        In order to keep batteries from consuming the power of DFT, it is turned off automatically 

unless operation is carried out within the preset time. However, when the measuring is carried out 
in immersion type thermocouple measuring mode (when timer is active), power to the DFT is   
automatically turned off unless DFT is operated within the preset time after timer action is over. 
 (Data on the PV display blinks.) 

      

• Auto-range

 

        When the PV indication changes from 999.9 to 1000 ( ), resolution automatically changes   

from 0.1 ( ) to 1 ( ) 

      

• Immersion type thermocouple measuring mode (EI)

 

        In the case of immersion type thermocouple measuring mode, if temperature exceeds the preset 

timer start temperature, the timer starts countdown, and the data is indicated on the PV display at the   
same time with the Time-up and the indication blinks. 

        (HOLD value indicates the value at the Time-up regardless of sensor burnout.) 

However, after blinking is cancelled or power is turned on, if the indicated temperature is higher   
than the value preset in timer start temperature setting, EI function does not work. 
To start EI function again, be sure to set the indication temperature -5 (10 ) or lower. 

      

• Decimal point place indication switch

 

        It is possible to set the use of decimal point for 999.9 ( ) or less. 
        However, if it is set not to use decimal point, resolution turns from 0.1 ( ) to 1 ( ) 

      

• Battery life warning indication

 

        When the battery voltage lowers, the battery life warning indication blinks and indicates that it is time 

to change batteries. 

      

• Input burnout indication

 

        [Overscale] 

        In the case of thermocouple and RTD input, “

” blinks on the PV display if the sensor is burnt   

out or the value exceeds the rated scale maximum value. 

(Immersion type thermocouple measuring mode [EI] does not have this function.) 
However, while Data HOLD, its value is indicated by priority. 

Thermo-hygro sensor (THD-700-P) input: 

In the case of temperature sensor (TD-S), if the sensor is burnt out or the value exceeds the rated   
scale maximum value, “

” blinks on the data display. 

In the case of humidity sensor (HD-S), if the value exceeds the rated scale maximum value, 
  “

” blinks on the PV display. 

When humidity sensor is burnt out, “

” blinks or “0.0” is indicated on the PV display. 

However, while Data HOLD, its value is indicated by priority. 

        [Underscale] 

In the case of thermocouple and RTD input, if the value goes below the rated scale minimum value 

” blinks on the PV display. 

However, while Data HOLD, its value is indicated by priority. 

        

Thermo-hygro sensor (THD-700-P) input: 

        In the case of the temperature sensor (TD-S), if the value goes below the rated scale minimum value 

” blinks on the Data display 

In the case of the humidity sensor (HD-S), if the value goes below the rated scale minimum value 
“0.0” is just indicated on the PV display. 

However, while Data HOLD, its value is indicated by priority. 
When sensor is burnt out in the immersion type thermocouple measuring mode (EI), “

” blinks 

on the Data display. 

      

• Temperature unit selection

 

        Temperature unit 

 

or

 

 can be selected. 

      

• Memory backup

 

        Backups the memory data when changing batteries 

      

• Memory data clearing

 

        If the power is turned on while the MEMO No. key is being pressed, all the data that is memorized   

reverts to the factory adjusted value “

”. 

Содержание DFT-700-M

Страница 1: ...gories Warning and Caution Depending on circumstances procedures indicated by Caution may have serious consequences so be sure to follow the directions for usage Procedures which may lead to dangerous conditions and cause death or serious injury if not carried out properly Procedures which may lead to dangerous conditions and cause superficial to medium injury or physical damage or may degrade or ...

Страница 2: ...odel name label is on the back of the case Fig 1 3 1 Do not drop DFT and sensors strike them with hard objects press hard on them as this may lead to the breakdown or malfunction The heat resistance of DFT sensor knob and cable section is low unlike sensor measuring part metal section so use them in a temperature ranging from 0 to 50 Do not use them in or near water which may cause breakdown or ma...

Страница 3: ...ing the MODE key for 1 second turns the measuring mode to the Peak HOLD P HOLD measuring mode and the display indicates P HOLD Pressing the MODE key again for 1 second reverts the controller mode to the measuring mode and P HOLD on the display goes out 12 LIGHT key Pressing the LIGHT key turns on the backlight which goes out automatically after 10 seconds 13 HOLD RST key In the measuring mode the ...

Страница 4: ...the battery case Insert as indicated by 3 and then swing them in 4 Fig 4 1 2 If batteries are inserted correctly DFT 700 is indicated on the LCD display for 5 seconds and then returns to the power switches OFF When removing the batteries After making sure that power is OFF push side of the batteries up in the direction indicated by 5 then swing them in the direction indicated by 6 and remove them ...

Страница 5: ...ermocouple can be used or not with the or key Not indicated when RTD Pt100 or Thermo hygro sensor is selected in the Sensor selection Selection item Normal measuring mode Immersion type thermocouple measuring mode Factory adjusted normal measuring mode Selects the Auto Power Off time with the or key Selection item Not used 10 min 30 min 60 min Factory adjusted 10 minutes Sets the measuring time af...

Страница 6: ... 700 P is measuring temperature and humidity P HOLD is indicated on the display and the maximum value of humidity data currently indicated and temperature data when humidity data is maximum are held Temperature is indicated on the Data display and humidity is indicated on the PV display When this function is selected the indicated values are not renewed as long as temperature and humidity do not e...

Страница 7: ...lue that was set in the timer start temperature setting EI function does not work To start EI function again be sure to let the indicated temperature be 5 10 or lower When P HOLD EI is indicated at the right bottom of the display This is Peak HOLD P HOLD measuring mode When Auto Power Off is set power is automatically turned off unless DFT is operated within a preset time To cancel Peak HOLD P HOL...

Страница 8: ...on turns from 0 1 to 1 Battery life warning indication When the battery voltage lowers the battery life warning indication blinks and indicates that it is time to change batteries Input burnout indication Overscale In the case of thermocouple and RTD input blinks on the PV display if the sensor is burnt out or the value exceeds the rated scale maximum value Immersion type thermocouple measuring mo...

Страница 9: ...CE 706 For measuring temperature of liquid and interior 400 PCE 707 For measuring general surface temperature 400 PCE 707L For measuring general surface temperature 400 PCE H7 Exclusive use tip for PCE 707 707L for exchange 400 PCE 709 For measuring temperature of liquid and interior 400 PCE 700 M For measuring surface temperature of mold etc Magnet type 300 7 2 RTD Pt100 Model Drawing Main object...

Страница 10: ... period 0 4 seconds Standard function Automatic cold junction temperature compensation Only thermocouple input Automatic POWER OFF Auto range EI Immersion type thermocouple measuring mode Decimal point indication switch Battery life warning indication Input burnout Overscale Underscale Temperature unit selection Multi range input Memory backup Memory data clearing Dust proof Drip proof IP65 Supply...

Страница 11: ...Cancel the Data HOLD D HOLD or Peak HOLD P HOLD Humidity sensor HD S of Thermo hygro sensor THD 700 P is out of order Replace the humidity sensor with a new one Indication of the PV display or Data display is abnormal or unstable Sensor input designation is incorrect Set the correct sensor input unit is incorrect Set the correct temperature unit AC is leaking from the measured object to thermocoup...

Страница 12: ... sure to turn the power OFF of DFT 700 M when DFT 700 thermo hygro sensor hereinafter THD 700 P is attached or detached from DFT 700 M Do not attach or detach THD 700 P in wet place or with wet hands because it may cause electric shock Warning The maximum temperature written in the table is the limit temperature of the measuring part Use such as grip cord etc in an ambient temperature 0 to 60 exce...

Страница 13: ...rica Ltd Reg Office Mail Address URL Toll Free 25 Whitefriars Drive Toronto Ontario Canada M3A 2L2 25 Whitefriars Drive Toronto Ontario Canada M3A 2L2 http www shinkona com 1 866 4SHINKO Tel Fax 416 444 0817 416 444 2361 ...

Отзывы: