27
10 Unit Converting
The function to convert units multiplies the weight of a sample on the balance by the saved
coefficient, and displays the result.
* Select [1 SEt 4] (Unit Converting) in Function 1.
1. Displaying the coefficient setting.
Make sure that the [#] sign is displayed and press
and hold the Function key for a few seconds.
(This operation may be performed with a container
placed on the balance.)
The display then blinks [C SEt].
The previously-saved reference sample weight
blinks.
2. Enter a coefficient.
Press the Zero/Tare key.
The digit furthest to the right blinks.
3. Changing
numbers.
Pressing the Zero/Tare key changes the
number as shown on the right.
4. Select digit places to specify a coefficient.
Pressing the Function key shifts the blinking digit to
the left, allowing you to set a higher-order digit.
After a number is set, the minus sign [M] blinks.
Pressing the key again makes the digit furthest to
the right blink.
5. Save
the
coefficient.
Press the Set key.*
(* Pressing the Print key cancels the setting of a
reference sample weight.)
6. Put a sample to weigh on the balance.
Put a sample to weigh on the balance.
The weight of the sample is multiplied by the
coefficient, and the result is displayed.
For example, Displayed value (250.0000)
= Coefficient (2.5) x Sample (100.0000 g)
Caution
The increment of the minimum readability is automatically set to 1, 2, or 5 depending on the
entered coefficient.
Continuous
press
Key released
Decimal point