10 Troubleshooting
FS series operation manual
-
80
-
Message
Cause
Countermeasures
L-Err
・
Sample weight is too light in comparison
with the memorized reference mass of the
percent scale.
t-Err
・
Addition computation was executed
doubly due to erroneous adding
operation.
・
Return the indication to zero,
confirm that an asterisk
「 」
mark
lights and then execute the
operation of addition computation.
・
Zero or minus addition computation was
executed on the plus side addition
computation.
・
When the indication is “0” or in the
negative state, addition computation
cannot be executed. Place an object
to be weighed before executing
addition computation.
・
Zero or plus addition computation was
executed on the minus side addition
computation.
・
When the indication is “0” or in the
positive state, addition computation
cannot be executed. Remove the
object to be weighed to make it
negative state before executing
addition computation.
Locked
・
In the state of being locked
・
Release the lock of a function
concerned from the setting menu.
(Refer to “7 Functions related to the
lock”.)
Message
Cause
Countermeasures
Err001
・
A system error
・
Take a note of the error number
and notify the store where you
purchased the product.
-
Err099
Message
Cause
Countermeasures
Err100
・
Communication error in the weighing
section
・
Check the scale cable connection.
Err101
Err102
Err103
Err104
Err112
・
Communication error in the power
supply box
・
Check the communication cable
connection.
Err113
Err114
Err120
・
Communication error
・
Notify the store where you
purchased the product.
Err121
Err122
Err123
Err124
Err200
・
Internal processing error
・
Notify the store where you
purchased the product.
Содержание FS100K1G
Страница 2: ...白紙 ...
Страница 95: ...FS series operation manual Appendix 85 Appendix 1 3 Dimensional outline drawing ...
Страница 96: ...Appendix FS series operation manual 86 Appendix 2 Operation of the setting menu Setting of various functions ...
Страница 97: ...FS series operation manual Appendix 87 Returns to the state of weighing after completion of setting ...
Страница 100: ...Appendix FS series operation manual 90 Hierarchy of the external input output functions can not be used ...
Страница 101: ...FS series operation manual Appendix 91 Hierarchy of functions related to the lock ...
Страница 102: ...Appendix FS series operation manual 92 Hierarchy of controlling and adjustment functions can not be used ...
Страница 106: ......
Страница 107: ......
Страница 108: ......