4 Function srelated to the performance
FS series operation manual
-
30
-
4-4
Stability discrimination frequency
Can not be used.
The larger numeric value is set, the higher stability is obtained.
1
Select a setting menu.
Select the stability discrimination frequency.
(“2-7-2 Operation of the setting menu, setting of
various functions” and “4-1 Hierarchy of functions
related to the performance”)
2
Set the stability discrimination frequency.
Inputting of the setting value
Push the [Direction] key.
Select
「
23.SC.
」
.
Input a setting value.
「
23.SC.0
」
:
16 times
(Severe)
「
23.SC.1
」
:
10 times
「
23.SC.2
」
:
8 times
「
23.SC.3
」
:
4 times
「
23.SC.4
」
:
2 times
「
23.SC.5
」
:
1 times
(Moderate)
4-5
Response speed
The larger numeric value is set, the higher stability is obtained.
1
Select a setting menu.
Select the response speed.
(“2-7-2 Operation of the setting menu, setting of
various functions” and “4-1 Hierarchy of functions
related to the performance”)
2
Set the response speed.
Inputting of the setting value
Push the [Direction] key.
Select
「
24.RE.
」
Input a setting value.
「
24.RE.1
」
:
1
(
Quick
)
「
24.RE.2
」
:
2
「
24.RE.3
」
:
3
「
24.RE.4
」
:
4
「
24.RE.5
」
:
5
(
Slow
)
Содержание FS100K1G
Страница 2: ...白紙 ...
Страница 95: ...FS series operation manual Appendix 85 Appendix 1 3 Dimensional outline drawing ...
Страница 96: ...Appendix FS series operation manual 86 Appendix 2 Operation of the setting menu Setting of various functions ...
Страница 97: ...FS series operation manual Appendix 87 Returns to the state of weighing after completion of setting ...
Страница 100: ...Appendix FS series operation manual 90 Hierarchy of the external input output functions can not be used ...
Страница 101: ...FS series operation manual Appendix 91 Hierarchy of functions related to the lock ...
Страница 102: ...Appendix FS series operation manual 92 Hierarchy of controlling and adjustment functions can not be used ...
Страница 106: ......
Страница 107: ......
Страница 108: ......