EinScan H V1.0.4.4-User Manual
-
May 2021
22
4.3.
How to do Calibration
Select Calibration to enter calibration interface. For first-time use, you will be guided to
STANDARD
CALIBRATION
directly.
1.
Put the calibration board vertically onto the paper position.
2.
Follow the instructions from the software to finish 5 positions.
3.
While holding the scanner, make sure it is upright to the center of the board. Bring the scanner
closer or farther to the board till all the required distance captured.
A pop-up of Calibration succeeded will appear. Click
Next
on the pop-up, software will take you to the
next step:
WHITE BALANCE
. Put the back side of the board on a stable and level table, and point the
scanner upright to the board. Move the scanner up and down till middle distance bar is highly bright
and scanner light projection ends. Hold the Scanner still when the Scanner is projecting light.
A pop-up of White balance calibration succeeded will appear. Click Next on the pop-up, you will enter
to scan mode.
Position 1
Position 2
Position 3
STANDARD CALIBRATION
Position 4
Position 5
White back side of the board
STANDARD CALIBRATION
WHITE BALANCE