Instruction Manual Appendix
23
7 Appendix
Appendix I Electromagnetic compatibility warning
The PCR Nucleic Acid Analyzer SWM-01 is in line with the emission and immunity requirements
specified in IEC 61326-1:2012 and IEC 61326-2-6:2012. The user has the responsibility to ensure the
electromagnetic compatibility environment to make the device work normally. Therefore, you can evaluate
the electromagnetic environment before the use of the device.
Table7-1
Electromagnetic Emissions
Emissions test
Compliance
Electromagnetic environment - guidance
Radiated emissions
CISPR11
Group 1
The PCR Nucleic Acid Analyzer SWM-01 RF energy only
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
Radiated emissions
CISPR11
Class B
The PCR Nucleic Acid Analyzer SWM-01 is suitable for
use in professional healthcare environments, but if used
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes, whatever additional measures are necessary.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions
IEC61000-3-3
Not applicable
Table7-2
Electromagnetic
Immunity
Test item
EMC basic
standards
Test value
Compliance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Air discharge: 2kV, 4kV, 8kV,
Contact discharge: 2kV, 4kV
A
Radiated RF
electromagnetic Fields
IEC 61000-4-3
3V/m, 80MHz-2.0GHz, 80%AM
A
Power-frequency
Magnetic Fields
IEC 61000-4-8
3V/m, 50Hz
A
Voltage Dips,
Interruptions, and
variations
IEC 61000-4-11
1 cycle 0%; 5 cycles 40%; 25 cycles
70%
5%, duration: 250 cycles
A
Electrical Fast
Transients and bursts
IEC 61000-4-4
1kV (5/50ns, 5KHz)
A