Instruction Manual Fault Diagnosis and treatment
22
slow or temperature control is
not accurate
Micro heater or fan is
damaged
Contact the supplier or manufacturer
Temperature sensor is
damaged
7
Unable to capture photos
Camera lens is damaged
Contact the supplier or manufacturer
Cable disconnection
The system pops up a
dialog box prompting
"Cycle duration with
detect
fluorescence
selected YES should be
over 20 seconds"
The cycle time of collecting the
fluorescence stages should set up
more than 20s
8
Location of image acquisition
is different
Camera lens shift
Contact the supplier or manufacturer
Chip card slot damage
9
Fluorescence detection value
is abnormal
Strong external light
source
Turn off the external light source
Photoelectric
system
damage
Contact the supplier or manufacturer
10
Timing error
Program error or clock
battery is exhausted
Contact the supplier or manufacturer
11
A pop-up dialog box prompts
"Please insert a USB flash
drive"
No USB flash drive is
inserted.
Insert the USB flash drive at the USB
interface
10
according
to
the
instructions manual.
12
A pop-up dialog box prompts
"Save failed"
U disk doesn’t have
enough storage space.
Clean up the historical data of the U
disk.
U disk is not compatible
with the instrument.
It is recommended to use the
matching USB flash drive or reformat
the USB flash drive before using.