![ShinewayTech SLS Seriesl OLT Series Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/shinewaytech/sls-seriesl-olt-series/sls-seriesl-olt-series_user-manual_1257999041.webp)
User’s Manual for Palm Optical Fiber Tester
Shineway Technologies, Inc.
- 35 -
Stabilized Laser Source
λ
ShinewayTech
®
uW
dBm
CW
MOD
dBm
AUTO OFF
nm
Optical Power Meter
λ
ShinewayTech
®
dB
dBm
Zero
uW
Ref
Set
dBm
AUTO OFF
nm
flange
Emitting optical jumper
Receiving optical jumper
On
Off
On
Off
F
IGURE
25.
S
KETCH MAP OF CONNECTION
2. Press the
[Ref]
key on OPM for several seconds until “HELD” appears on the LCD
screen, optical power meter will set the current measured dBm as reference value. The
display should read 0.00dB.
3. Connect both ends of optic fiber to be tested with laser source and power meter
through optical jumper, as in Figure26.