DPM-50 RF In Line Digital Power Meter User's Manual
Shineway Technologies, Inc.
- 30 -
[4]. VSWR
[5]. Return Loss
[6]. Crest Factor
[7].CCDF
Note: Equations:
[1]. Crest Factor (Logarithm value) = Peak Power (Logarithm) - Forward Power (Logarithm)
[2]. VSWR = (1 + Rho) / (1 - Rho)
[3]. Return Loss (Logarithm) = Forward Power (Logarithm) - Reflected Power (Logarithm)
Click
“Run Continues” to run a measurement. The interval and last time value can be
set in the
“Setting” menu. The measurement result will be displayed in the GUI by
two ways: number and dial. The number will be red while it is out of the setting range,
and be green while it is within the setting range. The data will be point by point added
in the historical trace. The user can click on the pages under the trace to check the
data trace for the different measurement items.
Curve chart
4) Info
The device information, Frequency, Setting are shown in
“Info” column.