![Shimadzu Sonialvision Safire Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/shimadzu/sonialvision-safire/sonialvision-safire_operation-manual_1256908029.webp)
29
M517-E077A
7 T r o u b l e s h o o t i n g
• Diagnostic table console
Message
Meaning
Position Detect Error XXX
P o s i t i o n d e t e c t e r r o r i n
operating parts
Position Error XXX
Potentiometer break detect
Motor Error XXX
Motor error
Procedure
Reset by pushing STOP
switch, or re-startup the
system. If the system is not
restored, contact Shimadzu
service representative.