7 T r o u b l e s h o o t i n g
• X-ray high voltage generator console
When the system detects a fault, an error message appears on X-ray high voltage
generator console.
Touch the
key to close the message screen, and then take action described
below for the error message.
Error messages
Messages related to the X-ray high voltage generator
(fault indicator) on the console illuminates and one of the following error messages
appears.
The system cannot expose while these messages are displayed.
Message
Meaning
Radio Over Current
Radiography tube current
e x c e e d e d s e t v a l u e
+ 200 mA.
Measured kV OVER
M e a s u r e d t u b e v o l t a g e
exceeded permitted range.
Starter ERR
S t a r t e r i s i n a b n o r m a l
situation or condition.
Starter is not working
Starter is not working when
fl uoroscopy.
I.F. OVER
Abnormal fi lament heating
current.
Line Voltage OVER
Supply voltage exceeded
permitted range.
Charge Volt ERR
Abnormal charging voltage of
primary smoothing capacitor.
Power Down
A b n o r m a l c o n t r o l c i r c u i t
supply voltage.
H. V. T Not Connected
A b n o r m a l c o n n e c t i o n t o
high-voltage transformer.
Arcing Trouble
Repeated arcing occurred.
Procedure
If this message appears
r e p e a t e d l y , c o n t a c t
S h i m a d z u s e r v i c e
representative.
Contact Shimadzu service
representative.