7 T r o u b l e s h o o t i n g
Error messages at power ON
Message
Meaning
Panel Battery WARNING
The data recording battery in
the operation panel must be
replaced soon.
Cabinet Battery WARNING
The data recording battery in
the control cabinet must be
replaced soon.
Panel Battery EMPTY
The data recording battery in
the operation panel is empty.
Cabinet Battery EMPTY
The data recording battery in
the control cabinet is empty.
X-ray switch ERR
T h e r a d i o g r a p h y b u t t o n
remains ON.
Fluo switch ERR
T h e fl u o r o f o o t s w i t c h
remains ON.
Procedure
Contact Shimadzu service
representative.
Messages related to Communication (Option)
(caution indicator) on the console illuminates and one of the following error
messages appears.
The system cannot expose while these messages are displayed.
Message
Meaning
Communication ERR
R e c e i v e d r a d i o g r a p h i c
conditions cannot be set for
the equipment.
Procedure
Change the setting at the
external instrument.
Other messages
(caution indicator) on the console illuminates and one of the following error
messages appears.
The system cannot expose while these messages are displayed.
Message
Meaning
Door/Interlock
The examination room door
is open.
AEC OVER
AEC failed and AEC backup
w a s u s e d d u r i n g A E C
radiography.
T h e fl u o r o s c o p y t a b l e i s n o t
prepared
The fl uoroscopy table is not
prepared due to incorrect
position of X-ray tube.
Procedure
Close the door.
I n c r e a s e r a d i o g r a p h y
tube voltage or extend the
radiography time.
C o n fi r m t h e d i a g n o s t i c
table confi guration.