Dirty cells and piping may cause phenomena such as an unstable
baseline or a large baseline noise due to pulsation of the solvent
delivery pump.
In this case, clean the flow lines with the following
procedures.
to
be
prepared
Syringe
1
I
Adapter
I
Water
I
Acetone
I
nitric
acid
I
I
I-
Mixing nitric acid with organic solvent may
form an explosive
substance. Be extremely careful when handling
acid.
Deliver acetone from the inlet port using the syringe and
adapter. During supply, turn the M o w
to deliver
solvent to the reference side.
(*)
The state in which
lamp is
lit
when
and
Use distilled water to flush the flow lines as in step
Flush the flow l i e s with
nitric acid.
Flush out the nitric acid with distilled water.
Replace with mobile phase. When the mobile phase used for
analysis is not miscible with water, replace the flow lines
with acetone and replace
it
with the mobile phase.
8-2
RID-1OA