5.2 Fundamentals
of
Operation
7.
Changing the range
(1)
When a recorder
is
connected
to the
RECORDER
terminal,
set the range at a value about 1.2 times as much
as the
expected maximum peak value. Maximum peak will be
about
80%
of the full scale ensuring that peaks remain
on-scale.
(2)
When
a
Chromatopac (integrator)
is connected to the
RECORDER terminal
terminal),
peaks
having
refractive index up to about
times as much
as the setting
range can be recorded. Set the measurement range at
a
value
about
1/80
of the expected maximum peak refractive index.
Normally, the recorder range
is
set
at
a value
of
about from
1
to 10 and the plot full scale is adjusted by changing
on
the Chromatopac side. The relationship
of
the setting
range and attenuation of Chromatopac
for the plot
full
scale
when connected to the RECORDER terminal
is
as
follows:
Plot
full
scale
Plot
scale
=
Setting range
x