The diagram shows flow lines in the
At
R
flow
OFF
When the
R
flow switch is
OFT,
the solenoid valve is open at side
A.
The solvent passing through the sample cell flows into the outlet
port, not into the reference cell.
At
R
flow
ON
When the
R
flow switch
is
ON,
the solenoid valve is open at side
B.
The solvent passing through the sample cell flows into the outlet
port after passing through the reference cell. This
is
used to fill the
sample cell and the reference cell with solvent of equal refractive
index before starting the measurement.
flow
ON
Outlet
port
port
Tubing
volume (At
R
flow
OFF)
Inlet port to flow cell
63.5
Flow cell volume
Flow cell to cell outlet port
280.2
S:
Sample
R:
Reference
1-6
RID-1OA