11
5) Loosen the clamp, E-20-3C (B10) mounted to the base unit ASSY (2), and remove the wire.
Fig. 2-20
6) Loosen the screw P4 M4 × 8 and flat washer M4 (2 locations) fixing the mount, analog board (B11) in place,
and remove the mount, analog board (B11).
7) Loosen the screw P3 M3 × 8 (4 screws) fixing the analog board ASSY (B3) in place, and remove the
analog board ASSY (B3).
8) Remove the connector (for the coil) inserted into the analog board ASSY (B3).
9) Use the above procedure in reverse to mount the new analog board ASSY (B3).
2.3.7. Replacing the Weight Loader ASSY (27)
1) Use the procedures in "2.2 Inspecting the Balance Interior" to remove as far as the unit cover (B5).
2) Loosen the screw P4 M4 × 8 (2 screws) and flat washer M5 (2 locations), remove the weight stopper (W3),
and remove the weight.
* When removing the weight (W4), be careful to avoid damaging the unit ASSY (1). (Be careful to
avoid letting the weight stopper (W3) rotating in tandem and coming into contact with the pan
supporter axis.)
* When performing re-assembly, be careful to avoid letting dust settle on the weight (W4).
3) Loosen the screw P4 M4 × 8 (4 screws), and remove the weight loader ASSY (27).
4) Use the above procedure in reverse to mount the new weight loader ASSY (27).
* Mount the weight loader ASSY (27) on cell width of the unit ASSY (1) in the balance left and right
directions.
* Be careful to avoid letting the weight loader mechanism come into contact with the lever.
* When mounting, press the weight loader ASSY (27) forward for the mounting position.
Fig. 2-21
Same interval
Press forward.
Clamp