7.6 Kinetic Method with Calibrator – Set-up Procedure
Select Method from Main Menu using >> then
↵
↵↵
↵
Select New from Method Menu using >> then
↵
↵↵
↵
Select Time Based from Mode Menu using >> then
↵
↵↵
↵
Select Kinetic from Mode Menu using >> then
↵
↵↵
↵
Select wavelength from list* using >> then
↵
↵↵
↵
Select Yes on Use Standard? Screen using >> then
↵
↵↵
↵
Set Standard value using >> then
↵
↵↵
↵
for each number column
Select required units using>> then
↵
↵↵
↵
for each unit’s column
Set Incubation time. Use >> then
↵
↵↵
↵
for each number column
Set measuring period. Use >> then
↵
↵↵
↵
for each number column
Note sample temperature if required using >> then
↵
↵↵
↵
(This facility only allows a record of the test temperature used to be printed
for traceability. The colorimeter does not control sample temperature).
Enter a method name using >> then
↵
↵↵
↵
for each letter
Select sample labelling format
You can print the method and check it before you save it
You can choose to edit, save or cancel your new method