
Sherwood Scientific Limited
260 91 000
Issue 1
26
th
May 2010
ECN 613
20
7.2 Concentration Method – Set-up Procedure
Select Method from Main Menu using >> then press
↵
↵↵
↵
Select New from Method Menu using >> then press
↵
↵↵
↵
Select Concentration from Mode Menu using >> then press
↵
↵↵
↵
Select wavelength from list* using >> then press
↵
↵↵
↵
Set Standard concentration value as outlined in the test kit. Use
>> then press
↵
↵↵
↵
for each number column
Set concentration units using >> then
↵
↵↵
↵
Note sample temperature if required using >> then
↵
↵↵
↵
(This facility only allows a record of the test temperature used to be printed
for traceability. The colorimeter does not control sample temperature).
Enter a method name using >> and then
↵
↵↵
↵
for each letter
Select sample labelling format
You can print the method and check it before you save it
You can choose to edit, save or cancel your new method