7.3 End Point Method with Factor – Set-up Procedure
Select Method from Main Menu using >> then press
↵
↵↵
↵
Select New from Method Menu using >> then press
↵
↵↵
↵
Select Time Based from Mode Menu using >> then press
↵
↵↵
↵
Select Endpoint from Mode Menu using >> then press
↵
↵↵
↵
Select wavelength from list*using >> then press
↵
↵↵
↵
Select No on Use Calibrator screen using >> then press
↵
↵↵
↵
Set Factor value using >> then
↵
↵↵
↵
for each number column
Set concentration units using >> then
↵
↵↵
↵
Set Incubation time using >> then
↵
↵↵
↵
for each number column
Note sample temperature if required using >> then
↵
↵↵
↵
(This facility only allows a record of the test temperature used to be printed
for traceability. The colorimeter does not control sample temperature).
Enter a method name using >> and then
↵
↵↵
↵
for each letter
Select sample labelling format
You can print the method and check it before you save it
You can choose to edit, save or cancel your new method