background image

2017/08/16                                                                                                                                                  Rev. A1 

 

18 

Category 

Systolic(mmHg) 

Diastolic(mmHg) 

Desirable 

120                    and 

80 

Normal   

120-129        and/or 

80-84 

Pre hypertension 

130-139        and/or 

85-89 

Hypertension: 

≥140            and/or 

≥90 

  Stage 1 Hypertension 

140-159      and/or 

90-99 

  Stage 2 Hypertension 

160-179      and/or 

100-109 

Hypertensive Crisis 

≥180            and/or 

≥110 

These categories were defined by the American Heart Association. This chart applies to adults 
age 20 and older. 

 

15.3. What is morning hypertension (morning surge)? 

Morning high blood pressure or morning surge is defined as the weekly average for morning 

blood pressure reading measured within 1 hour to 2 hours after awakening in the morning and 

exceeding 135/85mm Hg. Studies have shown that exaggerated morning blood pressure surge is a 

risk for cardiovascular events which includes ischemic and hemorrhagic stroke. Cardiovascular 

events have been shown to be exaggerated in the morning to coincide with morning high blood 

pressure. In fact heart attack, stroke and heart failure have been shown to fall particularly on a 

Monday amongst all the other days of the week. 

Organ damage and diabetic complications have also been shown to be linked with morning blood 

pressure surges just in the same way as small artery disease and multiple cerebral infarcts in elder 

members of society. Morning high blood pressure has shown some correlation with initial stage 

and progression of atherosclerosis. Patients with well controlled blood pressure may still have 

high morning blood pressure and this happens in 50% of the cases. Patients with morning 

hypertension have a 78% more chance of stroke compared with 48% of other hypertensive 

patients without morning high blood pressure. Morning hypertension has also been associated 

with changes in heart size and rhythm. This may lead to heart attack or heart failure. 

Morning Hypertension can only detect within 1 hour to 2 hours after awakening, recommended 

user monitor their own blood pressure at home.   

 
Reference Standard

 

·

IEC 60601-1:2005+A1:2012 Medical electrical equipment - Part1: General requirements for 

safety and essential performance. 

·

IEC 60601-1-2:2014 Medical electrical equipment - Part1-2: General requirements for basic 

safety and essential performance - Collateral standard: Electromagnetic compatibility – 

Requirements and tests 

·

IEC 80601-2-30:2009+A1:2013 Medical electrical equipment - Part 2-30: Particular 

requirements for the basic safety and essential performance of automated non-invasive 

sphygmomanometers. 

·

ANSI/AAMI/ISO 81060-2:2009 Non-invasive sphygmomanometers - Part 2:Clinical validation 

Содержание BF5010B

Страница 1: ...truction Manual BF5010B Thanks for purchasing the Arm Automatic Blood Pressure Monitor manufactured by Bpump Please read the instruction manual before usage Please dispose the Arm Automatic Blood Pressure Monitor and battery in accordance with local regulations ...

Страница 2: ... 10 8 1 Important Noted 10 8 2 Fitting the cuff 11 8 3 Body posture during measurement 11 8 4 Take Measurement 11 8 5 Mean value 13 8 6 Memory 14 9 Data transmission 15 10 Power off 15 11 Error Indication 15 12 Trouble Shooting 16 13 Specification 16 15 About Blood Pressure 17 15 1 What is Blood Pressure 17 15 2 What is high blood pressure 17 15 3 What is morning hypertension morning surge 18 Refe...

Страница 3: ...onment conditions specified in this manual Storage and transport environment Temperature 25 70 Humidity 10 95 Operating environment Temperature 5 40 Humidity 15 93 otherwise the device may not be able to achieve the claimed performance 9 Do not kink the connection tube or it may cause blood flow interference and result in harmful injury to the PATIENT due to continuous CUFF pressure 10 Do not use ...

Страница 4: ...erviced or maintained while in use with the patient 1 2 Contraindications 1 Do not use the blood pressure monitor for any other purpose except measuring the blood pressure of human body 2 Those who have arrhythmia diabetes blood circulation or apoplexy problem please use under the physician s instruction 3 Clinical testing has not been conducted on newborn infants and pregnant women Do not use in ...

Страница 5: ... unit or battery into fire Do not frequently run out of battery avoid affecting the battery life Do not forcibly boot after battery runs out avoid irreversible damage to the battery Use the correct adapter designed for your local voltage AC 100 240 V If the unit is to remain unused for an extended period more than 7 days please ensure that the battery retains a certain amount of electricity turn o...

Страница 6: ...f Label Fig 2 Note Symbol for TYPE BF APPLIED PART Display Fig 3 3 Power source 3 1 Battery 3 1 1 Installation and replacement 1 Remove battery cover 2 Load 3 standard AA alkaline battery as indicated in figure below Model BC ...

Страница 7: ... batteries will last for approximately 200 measurements if measurements are taken once a day at room temperature 23 C The batteries enclosed in the package are used for demonstration purpose It is possible these batteries will therefore not last for 200 measurements 3 2 Power Adapter Optional 1 Connect the AC adapter to the AC adapter connector of the main unit 2 Plug the socket into electric outl...

Страница 8: ...hut off the device 5 First touch mode Under the clock mode press the button to lit up the touch button area Fig6 press the button into next step 6 Standby mode Under the first touch mode without operation 2s to enter the standby mode which can automatically detect the current settings and display the Bluetooth symbol Horn symbol blood pressure unit and user group Fig7 Press the button or without o...

Страница 9: ...ne number Press the button to decrease one number Hold down the or button the number will change rapidly b Enter the digit of year number Fig8 c Press the Set button will proceed to month setting d Repeat step a to c to set month day hour and minutes Fig9 to Fig12 2 Unit with broadcast a Continues the broadcast setting press the or button will automatically turn on or off the broadcast system as s...

Страница 10: ... blood pressure if you are nervous or upset If we are nervous anxious or agitated our blood pressure will rise Rest for 5 10 minutes before taking a reading Sit in a comfortable relaxed position Don t move around or talk while taking the blood pressure Leave your legs in one position breath freely and calmly The blood pressure cuff should fit over about 3 4 of your upper arm It should easily go ar...

Страница 11: ...ccuracy may be affected Do not bend with the cuff or the air tube To detach the cuff unplug the air tube connector from the main unit Do not inflate prior fitting the cuff Change the cuff if it there is a leakage or if the cuff is not working properly Only use the manufacturer cuff to ensure the measurement accuracy 8 3 Body posture during measurement Relax place the elbow on the desk with palm fa...

Страница 12: ...atically deflate and the test result will be display on the screen The device will broadcast the result The WHO blood pressure level Optm Norm Pre H H1 H2 H3 will be displayed on the screen as show on Fig21 and Fig22 The blood pressure level classification and definition as show in Table 1 7 You may turn off the unit or compare with the previous results 8 If a problem occurs during the test Err wi...

Страница 13: ...rement or fail to perform as indicated press the button to discontinue measurement The time of the pressure reduced from 260mmHg 34 67kPa to 15mmHg 2kPa does not exceed 10s If cuff inflated up to 300 mmHg 40 kPa doesn t stop please remove the cuff or power off the unit 8 5 Mean value 1 Press the button to read the mean value 2 The screen will display the average data of the most recent 3 sets as s...

Страница 14: ...nd then turns to c c Display the systolic diastolic pulse rate and the blood pressure level as show on Fig26 Press the button to return a When reading the memory values if the result shows the heart mark it is suggested that the irregular pulse is detected Refer 7 1 to change user group Without any operation for 20s the unit will automatically turn to standby mode Under memory reading mode press d...

Страница 15: ...inflation rate is too low or main unit does not connect with the cuff 1 Reconnect the air tube connector with the main unit 2 Cuff or bladder leakage if necessary purchased a new one 3 Confirm the cuff is wrapped up correctly ref 8 retake the measurement Er 02 Weak Signal or cuff is too loose Heavy shock during the measurement Bad Signal moving or talking during the measurement Abnormal measuremen...

Страница 16: ...orrectly Measure value is different from the hospital or the value is inconsistent 1 Blood pressure value is varied during the day which also will affect by the human emotional and physical condition 2 Record the variance and consult to the doctor If the above suggestion doesn t remediable please consult the manufacturer 13 Specification Description Arm automatic blood pressure monitor Model BF501...

Страница 17: ...the walls of blood vessels and is one of the principal vital signs Two pressures are measured for a blood pressure reading Systolic blood pressure is a measure of blood pressure while the heart is beating Diastolic pressure is a measure of blood pressure while the heart is relaxed 15 2 What is high blood pressure High blood pressure also known as HBP or hypertension is a widely misunderstood medic...

Страница 18: ...surges just in the same way as small artery disease and multiple cerebral infarcts in elder members of society Morning high blood pressure has shown some correlation with initial stage and progression of atherosclerosis Patients with well controlled blood pressure may still have high morning blood pressure and this happens in 50 of the cases Patients with morning hypertension have a 78 more chance...

Страница 19: ...this device must accept any interference received including interference that may cause undesired operation FCC ID 2AAS7 BFA02 The device has been evaluated to meet general RF exposure requirement The device can be used in portable exposure condition without restriction Blood Pressure Measurement Chart Date Time SYS DIA Pulse Remark Date Time SYS DIA Pulse Remark Guarantee Card Product Model Produ...

Страница 20: ...hanged This product is medical device to ensure the accuracy of the product when using it we would like to continue to provide you with paid services after the guarantee periods Explanation of Symbols Symbol for batch code Symbol for Date of manufacture Symbol for manufacturer 0123 Symbol for CE Symbol for Follow operating instructions Symbol for TYPE BF APPLIED PART Symbol for electrical and elec...

Страница 21: ...eel Building Maqueling Estate Hi Tech Industrial Park Nanshan District 518057 Shenzhen PEOPLE S REPUBLIC OF CHINA Tel 86 755 26710795 26067119 Fax 86 755 26012025 Service Line 86 4006 755 009 E mail service bpump com cn Shanghai International Holding Corp GmbH Europe Address Eiffestraß e 80 20537 Hamburg GERMANY ...

Отзывы: