2017/08/16 Rev. A1
13
WHO Blood pressure level
Blood pressure range
Severe hypertension (H3)
Systolic≥180mmHg or/and Diastolic≥110mmHg
Moderate hypertension (H2)
Systolic≥160mmHg or/and Diastolic≥100mmHg
Mild hypertension (H1)
Systolic≥140mmHg or/and Diastolic≥90mmHg
Prehypertension
(
Pre.H
)
Systolic≥130mmHg or/and Diastolic≥85mmHg
Normal (Norm)
Systolic≥120mmHg or/and Diastolic≥80mmHg
Optimal (Optm)
Systolic <120mmHg and Diastolic <80mmHg
Notes:
Do not self-diagnosis according to measurement results. Follow the instructions of your
physician or licensed healthcare provider.
If the device caused any discomfort during measurement or fail to perform as indicated, press
the
【 】
button to discontinue measurement.
The time of the pressure reduced from 260mmHg (34.67kPa) to 15mmHg (2kPa) does not
exceed 10s.
If cuff inflated up to 300 mmHg (40 kPa) doesn’t stop, please remove the cuff or power off
the unit.
8.5. Mean value
1)
Press the
【
】
button to read the mean value.
2)
The screen will display the average data of the most recent 3 sets as show on Fig23 and
broadcast the data.
3)
Press the
【
User
】
button to change user group.
When reading the mean value, press the
【
】
button,the average data will flick
and the device broadcasts the data again.
Without any operation for 20s, the unit will automatically turn to standby mode.
Under mean value mode, press down the
【
】
button for 3s to return the clock
mode, press and hold the
【
】
button exceed 6s to shut off the unit.
Fig23