INSTALLATION
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the unit, verify each of the following requirements is satisfied.
Power Source
: The wall power outlet must meet the power requirements listed on the unit data
plate.
Model
AC Voltage
Amperage
Frequency
SVAC1
110 – 120
7.0
50/60 Hz
SVAC2
110 – 120
10.0
50/60 Hz
SVAC4
110 – 120
13.0
50/60 Hz
•
Supplied voltage must not vary more than 10% from the data plate rating
. Damage to the
unit may result if the supplied voltage varies more than 10%.
•
The wall power source must be protective earth grounded.
•
Use a separate circuit to prevent loss of the unit due to overloading or circuit failure.
•
The recommended wall circuit breakers for these units are 15 amps.
•
The wall power source must conform to all national and local electrical codes.
Power Cord
The unit must be positioned so that all end-users can quickly unplug the oven in the event of an
emergency.
•
Each unit comes provided with a
125-volt, 15-amp, 9ft 5 in (2.86 m) NEMA 5-15P
power
cord.
•
Always use this cord or an identical replacement.
Fuses
Each unit comes with a fuse installed in a fuse holder immediately adjacent to the power cord inlet.
•
The fuse must be installed and intact for the unit to operate.
•
Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
•
Fuse type:
o
T16A 250V 5x20mm
Содержание SHEL LAB SVAC1
Страница 1: ...Installation Operation Manual SVAC1 SVAC2 SVAC4 VACUUM OVENS 110 120 Volts...
Страница 6: ...6 P a g e TABLE OF CONTENTS...
Страница 12: ...12 P a g e INTRODUCTION...
Страница 24: ...24 P a g e RECEIVING...
Страница 34: ...34 P a g e SYMBOLS...
Страница 74: ...74 P a g e UNIT SPECIFICATIONS...
Страница 77: ...77 P a g e PARTS LIST...