28 |
P a g e
CONTROL PANEL OVERVIEW
Set CO
2
Display and Controls
Shows the current concentration of CO
2
as a percentage of the incubation chamber
atmosphere. The display has a range of OFF (0%) to 20% and an accuracy of 0.1%. The
display shows “LO” until the CO
2
sensor registers a concentration in the chamber greater
than 0%. A few minutes of CO
2
injections may be required when initially starting the unit.
The
Up
and
Down
arrow buttons are used to access the CO
2
Setpoint (SP) or Calibration
Offset (C O) display modes and input the CO
2
setpoint or calibration adjustment value.
Red LED alarm indicators marked High and Low illuminate whenever the CO
2
concentration deviates
by ±1% or greater from the setpoint. The yellow LED marked Mute illuminates whenever an audible
deviation alarm is being muted. See the
entry on page 38 for more
information.
The green CO
2
Injecting light illuminates when the incubator is injecting CO
2
into the incubation
chamber. Injections are accompanied by a clicking sound that is the CO
2
solenoid opening and
closing.
Water Jacket Low
The red Water Jacket Low light illuminates when the water level of the jacket has fallen to a level
that impacts the incubator temperature uniformity and stability. Add water to the jacket as soon as
this light illuminates. Fill until the light turns off. See the
entry on page 23 for
instructions.