INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
These units are built for use indoors at room temperatures between
15
°
C and 30
°
C (59
°
F and 86
°
F)
, at
no greater than
80% Relative Humidity
(at 25
°
C / 77
°
F The ambient temperature should not change by
2°C (3.6°F) or more during operation.
Operating outside these conditions may affect the incubator temperature performance.
When selecting a location to install the unit, consider all environmental conditions that can adversely
impact its temperature performance. These include:
•
Proximity to ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling vents or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight
R
EQUIRED
C
LEARANCES
These clearances are required to provide air flows for ventilation and cooling.
4 inches (102 mm)
of clearance is required on the sides and back.
2 inches (51 mm)
of headspace clearance is required between the top of the unit and any overhead
partitions.
Power Cord
2” (51 mm)
4” (102 mm)
4” (102 mm)