The SDHmini will display a moving bar until it assesses that
the sensor has stabilised (up to 10 minutes).
The full scale value, 10 °C dewpoint, is now displayed.
Press when complete.
Close the desiccant head when instructed and press .
Pressing during
CalSet
aborts the process and the value
present at the start of the aborted
CalSet
process is reinserted.
Press to continue/accept changes.
The SDHmini will return to the normal
operational mode.
5.3 Calibration Procedure Error Messages
If the instrument fails calibration, a message will be displayed on the screen until
acknowledged by the user, e.g.
AutoCal Failed
Sensor Open Circuit
or
AutoCal Failed
Sensor Short Circuit
This will mean replacing the sensor or its interconnecting cable in some cases.
6 Sensor Related Settings
On the
Configuration
screens there exists a collection of measurement settings
associated with the sensors and its environment.
6.1 Sensor Type
Here you may view the type of sensor
installed in the SDHmini and its range.
6.2 Selecting Gas Types
The SDHmini default setting is for ideal gas behaviour, however, there is an option
to change the setting
to natural gas which
uses the appropriate
calculation.