![Shaw SDHmini Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/shaw/sdhmini/sdhmini_instruction-manual_1255076012.webp)
3.9 Pressure and Flow Control
In most cases the gas to be sampled will be at a positive pressure, which will need
to be reduced to atmospheric pressure. At low pressures, up to around 8 barg, this
can be done by including a simple needle valve (5) to control the flow and thereby
the pressure dropping across it. Ideally this would be the isolation valve (1) on the
sample point, thus keeping the number of joints and glands, which are points of
potential moisture ingress to a minimum.
At higher pressures it is advisable to reduce the pressure across a pressure
regulator (4). Any regulator used should be stainless steel and of a low swept
volume. Typical laboratory or welding type regulators are not suitable because
they have elastomeric diaphragms which are permeable to water vapour and the
high swept volume will be slow to purge.
3.10 Particulate Contamination
If the sample gas is likely to contain particulate material this should be kept away
from the sensor by including a small inline particle filter (2) in the sample system.
If the surface of the sensor becomes partly coated with particles it will not affect
the calibration of the sensor but will restrict the movement of water molecules
across the surface so increasing the time it takes for the sensor to come to
equilibrium with the water vapour concentration around it. If the solid material is
hygroscopic, such as desiccant dust, whenever there is a change in the moisture
concentration in the sample, this material must come to equilibrium with the
moisture content before the sensor can respond to the change, which will cause
delays in obtaining accurate results.
3.11 Liquid Contamination
If the sample gas may contain liquid contamination, such as oil mist, this should
be kept away from the sensor by including a coalescing filter or an appropriate
membrane filter (2) in the sample system.
If the surface of the sensor becomes partly coated with liquid, again, it will
not affect the calibration of the sensor. It will restrict the movement of water
molecules across the surface so increasing the time it takes for the sensor to
come to equilibrium with the water vapour concentration around it. If the liquid is
hygroscopic, such as glycol, the buffering effect described for desiccant dust may
be so great that the sensor becomes too slow to use.
12
Содержание SDHmini
Страница 1: ...Model SDHmini Dewpoint Meter Instruction Manual 3006 SDHmini User Manual Iss 1 1...
Страница 2: ......