- 15 -
SI 510 WEIGHING INDICATOR
Max decimal point will be 0.0001, and digit can be selected among 1,2,5,10,20,50.
Digit and decimal point must be fulfilled under the below condition.
In this step, If there is unstable condition such as some forces or vibration on the
scale part, "Error-A" will be displayed, and "DEAD value" will not be calculated.
Please remove the cause of the force or vibration and process it again.
STEP 3 – Decimal point and division setting
1) After "DIVI" is displayed, locate the
decimal point with "F1"key.
2) Set the division with "F2" key.
Press enter key to save.
STEP 4 – Measuring the "DEAD" Weight of Weighing Scale
1) When "DEAD" displays, Press Enter key.
2) The indicator will calculate dead weight of
scale part automatically. (While this process,
there should be nothing on the scale part.)
Indicator will search "DEAD weight" during
10secs automatically to find the best
condition.