SI 510 WEIGHING INDICATOR
6-1-9. Modbus Memory Map
-
If you need more detailed Modbus Resister Map, please refer to the
information on the home page or contact our support team.
-
RO : Read Only
- RW : Read Write
- Each P/N’s set point can’t over max capacity of Indicator.
ex)35.00kg = 3,500 (0xDAC)
- When you input date and time, it should be 6digit.
ex) 1
st
January 2014 = 140101 (0x22345)
15(H) : 50(M) : 17(S) = 155017 (0x25D89)
- Refer the memory register for regarding Lamp, Error, Digital Input, Standard Key,
Special Key
- Modbus Function Codes
'03' (0x03) : Read Holding Registers
'04' (0x04) : Read Input Registers
'06' (0x06) : Write Single Registers
'16' (0x10) : Write Multiple Registers
- CRC Check Method is CRC-16.