Purpose
The multiple test set SPE 2 can be used to check the function
control and the adjustments to gas detection, -warn and
-measuring instruments.
The following functions can be carried out:
Testing of gas detection devices with ppm incicator,
Testing of gas detection-, gas warn-, and gas
measuring devices with LCD- or vol.% indicator,
Measuring underpressure and flow of pumps.
1.2
Contents of Set
The test set SPE 2 comprises the following:
2 adapter hoses for concentrations of 10 ppm
methane CH
4
,
2 adapter hoses for concentrations larger than
10 ppm methane CH
4
,
1 adapter hose for outside air.
Please mark both hoses meant for 10 ppm CH
4
, so that they are
not being mixed up with those hoses meant for high concentrations.
The multiple test set comes without test gases.
1.3
Accessories
We offer the following test gas as 0,4 l or 2,0 l gas pressure bottles
with a fill pressure of max. 120 bar:
10 ppm CH
4
,
synthetic air,
1.00 vol.% CH
4
,
140 ppm CO or
2.20 vol.% CH
4
,
17.50 vol.% O
2
.