background image

9

© Setra Systems, Inc. All Rights Reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.

8.0 RH and T adjustment of the probes

8.1 Definitions

 

Calibration

 

The calibration documents the accuracy of a measurement 

device. The device under test (specimen) is compared 

with the reference and the deviations are documented in a 

calibration certificate. During the calibration, the specimen 

is not changed or improved in any way.

Adjustment

The adjustment improves the measurement accuracy of a 

device. The specimen is compared with the reference and 

brought in line with it. An adjustment can be followed by a 

calibration which documents the accuracy of the adjusted 

specimen.

8.2 Sensing adjustment with SRH400 basic 

device

•  1-point RH or T adjustment for optimal performance over 

a narrow RH or T range.  

 

Example: climate monitoring in a pharma manufacturing 

site. RH range: 40…60 % RH, RH adjustment point: 50 % 

RH T range: 15…25 °C (59…77 °F)  T adjustment point: 20 

°C (68 °F)

•  2-point RH or T adjustment for accurate measurement 

over a wide RH or T range. 

 

Example: outdoor measurement  

RH range: 10 … 100 % RH, RH adjustment points: 40 

% RH, 70 % RH T range: -30 … 35 °C (-22...95 °F), T 

adjustment points: -10 °C (14 °F), 15 °C (59 °F)

•  The adjustment data is saved into the probes. The SRH400 

basic unit is not affected by the adjustment procedure. 

Consequently, it is possible to replace the probes at any 

moment.  

•  Start the adjustment procedure with the low point (e.g. 30 

% RH) and continue with the high point (e.g. 80 % RH)

•  For accurate calibration or adjustment, the specimen and 

the reference must have same T. 

•  During calibration or adjustment the T shall remain 

constant. 

•  Allow min. 30 min, stabilization time for RH calibration or 

adjustment. 

•  If needed, clean the sensors and replace the filter cap by 

a new one. 

8.2.1 2-point RH and T adjustment

 

1. Set the jumper to RH calib or to T Calib.  

2. Place the probe into the corresponding calibrator and let 

it stabilize at the first point (low) for at least 30 minutes. 

3. Press and hold S2 for 3 seconds. The LED “Calib” will light 

and the symbol “CAL<” will appear on the SRH400 display. 

4. Push S1 (up) or S2 (down) to bring the measured 

value in line with the reference in steps of 0.1. The 

actual corrected value is indicated on the display 

and can be measured at the analogue output. 

5. Press and hold S1 for 3 seconds to store the value into the 

probe and end the first (low) point adjustment. The LED 

“Calib” will be deactivated and the display will show “CAL”

6. Alternative: Press and hold S2 for 3 seconds to exit 

the adjustment mode without storing the adjustment 

data. The LED “Calib” will be deactivated and the 

symbol “CAL” disappears from the LC display. 

7. Set the calibrator for the second (high) adjustment 

point and let the probe stabilize for at least 30 min. 

8. Press and hold S1 for 3 seconds. The LED “Calib” will light 

and the symbol “CAL>” will appear on the SRH400 display. 

9. Push S1 (up) or S2 (down) to bring the measured value 

in line with the reference in steps of 0.1. The actual 

corrected value is indicated on the display and can be 

measured at the analogue output. Press and hold S1 for 

3 seconds to store adjusted value in the probe and end 

the second (high) point adjustment. The LED “Calib” 

will be deactivated and the display will show “CAL”. 

 

Alternative: Press and hold S2 for 3 seconds to 

exit the adjustment mode without storing the 

adjustment data. The LED “Calib” will be deactivate 

and the symbol “CAL” disappears from the LC display.

 

Содержание SRH400

Страница 1: ...Setra Systems Inc 159 Swanson Road Boxborough MA 01719 800 257 3872 www setra com Operating Instructions Model SRH400...

Страница 2: ...t is for information purposes only and is subject to change without notice Setra Systems assumes no responsibility for any errors or for consequential damages that may result from the use or misrepres...

Страница 3: ...8 0 RH and T adjustment of probes 8 1 Definitions 8 2 S07 adjustment with SRH400 basic device 8 3 1 point RH and T adjustment 8 4 Return to factory calibration 8 5 Loop adjustment and calibration in...

Страница 4: ...led in a manner that does not have a negative effect on its safe use The user is responsible for observing all applicable safety guidelines local and international with respect to safe installation an...

Страница 5: ...as on the optional display One or two point adjustment for RH and T of the transmitter canbeeasilyperformedwithpushbuttonsontheelectronics boardoftheSRH400basicunit Alternatively theprobescanbe adjus...

Страница 6: ...H400 version with remote probe please see below 4 4 probes connected to SRH400 with extension cables SRH400 with remote sensing probe The probes can be also connected to the SRH400 housing with extens...

Страница 7: ...e electrically isolated from the GND Locate the SRH400 the probe cable and the output cables as far as possible from sources of electromagnetic disturbances For the analogue outputs use either shielde...

Страница 8: ...on indicates adjustment mode 1 flash confirms reset to factory calibration 4 S1 Functions 1 point RH or T adjustment RH 50 RH T within the upper half of the T scale 2 point RH or T adjustment high ca...

Страница 9: ...hall remain constant Allow min 30 min stabilization time for RH calibration or adjustment If needed clean the sensors and replace the filter cap by a new one 8 2 1 2 point RH and T adjustment 1 Set th...

Страница 10: ...S2 for 5 seconds The return to factory calibration is indicated by one flash of the LED Calib 8 5 Loop adjustment and calibration in the field Loop calibration or adjustment in the field as required b...

Страница 11: ...15 35 V DC 15 29 V AC power supply 20 35 V DC RL 500 11 35 V DC RL 50 SRH400 Analog volts For the SRH400 with current output an ammeter shall be connected as in Fig 6 power supply for 0 1 V 10 35 V D...

Страница 12: ...tput Iout mA 16 Ref Tmax 4 mA For voltage output Uout V Abb Ref Tmax Where Abb voltage scaling 1 V 5 V 10 V Ref reference value of temperature 45 C 113 F resp 5 C 41 F Tmax max temperature scaling e g...

Страница 13: ...ers HA010103 SRHSS Stainless steel sintered filter Filters HA010104 SRHPG Plastic grid filter Filters HA010105 SRHPF PTFE filter 12mm Filters HA010106 SRHMG Metal grid filter Filters HA010115 SRHHP H2...

Страница 14: ...To assure prompt handling please make sure the RMA number is on the outside of the box and a copy of the service request from is included in the shipment If applicable include a copy of the PO for re...

Страница 15: ...s notified in advance of and the product is returned to SETRA transportation prepaid Unless otherwise specified in a manual or warranty card or agreed to in writing and signed by a SETRA officer SETRA...

Страница 16: ...Setra Systems Inc 159 Swanson Road Boxborough MA 01719 800 257 3872 www setra com SS SRH400 Rev B 09 2018...

Отзывы: