112 DF-760E
User Interface
8.000 VDC. This process can be repeated as often as needed. When done, the
NEXT
key is
used to move to the
Done
field and the
Enter
key is hit to leave the screen. The condition
of the analog output before the test is restored when the test is concluded. See the section
on analog outputs found on page 42 for additional information.
Figure 129: Test Moisture Analog Output
7.7.8.2 Test Oxygen Analog Output
The Test Output screen, as shown in Figure 130 is used to calibrate the oxygen analog
recorder output. When the Test Output screen is selected, the
NEXT
key is used to move
to the percentage field where the arrow keys (
and
) are used to choose the
appropriate setting. The
NEXT
key is then used to move to the
Apply
field where the
Enter
key is hit to set the analog output to the selected value. The analog output response
should match the value that was entered. For example, if 80 percent is entered for the
percent full scale level, and the analog output is set for 0 to 10 VDC, the analog output is
8.000 VDC. This process can be repeated as often as needed. When done, the
NEXT
key is
used to move to the
Done
field and the
Enter
key is hit to leave the screen. The condition
of the analog output before the test is restored when the test is concluded. See the section
on analog outputs found on page 42 for additional information.
Figure 130: Test Oxygen Analog Output
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...