42 DF-760E
Connecting to External Devices
See the chapter on Communications on page 113 for additional information on setting unit
ID’s and baud rates.
A program to facilitate serial communications is available from Delta F.
Pin #
Signal
Description
J5-8
GND
Ground
J5-7
TX-
4 wired 485 paired with TX+
J5-6
Key
Unused
J5-5
RX-
4 wire 485 paired with RX+
J5-4
Unused
J5-3
Unused
J5-2
TX+
Data transmitted by the analyzer via RS-232 or RS-485
J5-1
RX+
Data received by the analyzer via RS-232 or RS-485
Table 2: Pin-out of Serial Comm Connector J5
Pin assignments
DF-760/750
Connector J5
PC-DB9/RS-232
PC-DB25
/
RS-232
PC/RS-485
converter
1
: RX+
3
: TD
2
: TD
TX+
2
: TX+
2
: RD
3
: RD
RX+
8
: Gnd
5
: Gnd
7
: Gnd
5:
RX-
TX-
7:
TX-
RX-
Table 3: Serial Communications Connections
6.2
Analog Signal Outputs – J4, J10
Two sets of analog voltage outputs correlating to the front panel display readings are
provided on the rear of the analyzer. The analog voltage output for the oxygen signal is
provided through connector J10. The analog voltage output for the moisture signal is
provided through connector J4. The full scale voltage for each is set at the factory at the
time of order to: 0 to 1 VDC, 0 to 2 VDC, 0 to 5 VDC, or 0 to 10 VDC. These outputs are
electrically isolated from all other analyzer outputs, and from chassis (Earth) ground. See
page 65 and page 97 for additional information on setting the Moisture and Oxygen
Analog Outputs. These outputs may be tested with the use of the analog voltage test
routine found on page 111.
6.3
4-20 mA Outputs – J4, J10
Two sets of 4-20 mA analog outputs correlating to the front panel display readings are
provided on the rear of the analyzer. The 4-20 mA output for the oxygen signal is provided
through connector J10. The 4-20 mA output for moisture is provided through connector
J4. These outputs are electrically isolated from all other analyzer outputs, and from chassis
(Earth) ground.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...