Calculate Threshold, ReSpan
and
Start
buttons are only accessible
when calibration is enabled.
Other parameters displayed on the grid are:
Area Count
: this is the area that covers the peak in counts (digital converter
units). This area is used to generate the concentration displayed in the Result
column.
Response factor
: this is the area counts divided by the value in the Result
column.
Automatic
Re-Span:
when the box
is checked, the analyser automatically
performs a re-span after a cycle of calibration reaches an end (the same action
as pressing the
Re-Span
button). This means that after pressing the
Start
button in
Span
calibration, the cycle will start and then end, and the analyser
will automatically calculate the value of this calibration using the last measured
area counts and displays it in the
Result
column.
1.10 Calibration of the 4-20mA outputs
Click
Calibration
on the menu bar and then
4-20mA outputs
to access the
CALIBRATION-4-20MA OUTPUTS
menu
(see Figure 12).
Figure 12 : Calibration-4-20 mA outputs menu