a. Wait at least 20 minutes after eating or drinking before testing. Alcohol remaining in
the mouth, or even excess saliva, may interfere with testing. This includes
alcohol-based mouthwash or similar products. This is the most common cause of
improper use and incorrect result.
b. Avoid testing in high wind or enclosed spaces. Poor ventilation may also lengthen the
time required for warm up between tests.
c. Do not blow smoke, saliva or other contaminants into the mouthpiece, as the sensor
may become damaged.
d. Do NOT use iSOBER 30 as a tool to drink and drive.
e. Keep out of extreme temperatures and out of the reach of children.
f. To maintain accuracy, professional of the iSOBER 30 must be performed every 1000
tests or at least once a year, whichever comes first. To recalibrate, you must mail your
device to the address at the end of this manual. Please contact us for details and
related fees.
Warnings
Accuracy
Indication of B.A.C
Warm up time
Response time
Mouthpiece
Power supply
Weight
Dimensions(Inches)
Temperature conditions
Calibration interval
±5% at 0.050 %BAC
0.000 to 0.400%BAC by FND Display
3 seconds~4 minutes
within 10 seconds at 0.100%BAC
Provided Individual packed mouthpiece
Two 1.5V “AAA” Alkaline batteries only
2.33 oz (Including batteries)
3.94(H) X 1.73(W) X 0.59(T)
Operation 41°F to 104°F /
Storage 32°F to 122°F
12months or 1,000 tests(Whichever comes first)
Specifications