background image

Warning Display

If you see “Flo” in the display, the breath sample was not strong enough 

for measurement. Please power on the unit again and make sure the 

breath is strong and steady during sampling.
If you see “bAt” in the display, the battery power is low.

Please replace the batteries.

Unit is on grace period of calibration, please contact local 

distributor for the recalibration

Out of operating temperature, 

please follow operating temperature condition

Out of required calibration period, please contact local distributor 

Problem with sensor, please contact local distributor 

iSOBER30 memorizes various information related with test counting and time.

Especially the warning of remaining counts & dates until next calibration is helpful to 

prepare calibration service on time.

To enter the menu mode, hold power button 5 seconds to turn on the device. The latest 

test result will be shown briefly and then 3 selectable menu                                    will be 

displayed. 

To move between 3 menus, press power button once.

To confirm target menu, press & hold power button 3 seconds

To out of menu mode, select             or press & hold power button 3 seconds under the 

toggle displaying is going on

Menu mode

Содержание iSOBER 30

Страница 1: ...to 0 40 BAC It is easy to operation and handy compact size of breath alcohol tester and high accuracy with various functions will make a special experience Components Clear RED FND type Mouthpiece slot BAC of measurement unit One power button operation 2AAAAlkaline Batteries compartment Breath outlet hole Do Not block while using it 3 decimal point display e g Displaying 010 means 0 010 BAC ...

Страница 2: ... days counting remaining until a next calibration Breathalyzer Mouthpieces x5 AAA Batteries x2 Zipper case and User s manual Contents 1 Press power button 2 seconds then battery level will show as Full Middle Low 2 Wait until you can see 0 sign Warming up is going on 3 When 0 appeared on the display blowing through the mouthpiece into the unit for 4 seconds until hearing click crack sound 4 The re...

Страница 3: ... and out of the reach of children f To maintain accuracy professional of the iSOBER 30 must be performed every 1000 tests or at least once a year whichever comes first To recalibrate you must mail your device to the address at the end of this manual Please contact us for details and related fees Warnings Accuracy Indication of B A C Warm up time Response time Mouthpiece Power supply Weight Dimensi...

Страница 4: ...on period please contact local distributor Problem with sensor please contact local distributor iSOBER30 memorizes various information related with test counting and time Especially the warning of remaining counts dates until next calibration is helpful to prepare calibration service on time To enter the menu mode hold power button 5 seconds to turn on the device The latest test result will be sho...

Страница 5: ... time record Toggle displaying Toggle displaying It means The latest result was 0 020 BAC in 3 hours ago Ready to test It means The 10th latest result was 0 017 BAC in 2 days ago Press button once The numbers on toggle displaying below chart are examples to help user understanding ...

Страница 6: ...laying Toggle displaying Toggle displaying It means Unit s tested 5 162 times after calibration It means Unit s tested 5 162 times in total It means Unit s detected positive alcohol 5 162 times in total It means Unit will be required calibration service in 53 days or 991 tests whichever comes first Dot on the display is in thousands Ready to test Ready to test ...

Страница 7: ...eory Any act or failure to act based on a reading from this product shall be at the test subject s or user s own risk Sentech USA INC assumes no responsibility for consequences to or of test subjects who use this product and later are shown to have been under the influence of alcohol or to have had their judgment or any mental or bodily function impaired by alcohol Correlation between breath alcoh...

Страница 8: ...maining terms shall not be affected or impaired EXCEPTAS PROVIDED INTHIS WARRANTYANDTOTHE MAXIMUM EXTENTPERMITTED BYLAW SENTECH USAINC IS NOT RESPONSIBLE FOR DIRECT SPECIAL INCIDENTAL OR CONSEQUEN TIAL DAMAGES UNDER ANY LEGAL THEORY INCLUDING BUT NOT LIMITED TO LOSS OF USE LOSS OF REVENUE OR INCOME LOSS OF ACTUAL OR ANTICIPATED PROFITS INCLUDING LOSS OF PROFITS FROM A CONTRACT LOSS OF THE USE OF M...