RS485 COMMUNICATION OUTPUT TYPE 

1. Interface description 

Wide voltage power supply input 10-30V DC power supply. When connecting 485 signal line, it is noted that the two 
lines a and B cannot be connected reversely, and the address between multiple devices on the bus cannot conflict.

Line color

Explain

Power

brown

Positive power supply (10 ~ 30V DC)

black

Power supply negative (GND)

Output

yellow

485-A

blue

485-B

2. Installation and use of configuration software 

2.1. Software selection 

Open  485  data  package,  select  "debugging  software"  -  "485  parameter  configuration  software",  find 

(Parameter 

configuration tool.exe) 

and open it. 

2.2. Equipment connection

2.3. Parameter setting 

(1)  Select  the  correct  COM  port  (check  the  COM  port  in  my  computer  -  properties  -  Device  Manager  -  port).  The 
following figure lists the driver names of several different 485 converters.

3. Computing method 

3.1 conversion calculation of current mode output signal 

The  measuring  range  is  0  ~  40m  /  s,  4  ~  20mA  output.  When  the  output  signal  is  12mA,  the  current  wind  speed  is 
calculated. The span of wind speed range is 40m / s, which is expressed by 16mA current signal. 40m / S / 16mA = 
2.5m/s/ma, that is, the current changes 1mA, and the wind speed changes 2.5m/s. Then the measured value can be 
calculated. The measured value 12ma-4ma = 8Ma. 8Ma * 2.5m/s/ma = 20m / S, and the current wind speed = 20m / s. 

3.2 calculation of voltage mode output signal conversion 

The measuring range is 0 ~ 40m / s. take 0-10V output as an example, when the output signal is 5V, the current wind 
speed is calculated. The span of wind speed range is 40m / s, which is expressed by 10V voltage signal. 40m / S / 10V 
= 4m / S / V, that is, every 1V change of voltage corresponds to 4m / s change of wind speed. The measured value is 
5v-0v = 5V. 5V*4/m/s/V=20m/s

 The current wind speed is 20m / s.

Содержание WS303U

Страница 1: ... WS303U Ultrasonic Wind Speed Direction Sensor Product Manual ...

Страница 2: ...sensor can measure the instantaneous value of wind speed and direction at the same time It is widely used in the fields of meteorology ocean environment airport port laboratory industry agriculture and transportation The equipment with built in electronic compass no longer has the requirement of orientation during installation It only needs to ensure the horizontal installation The whole machine s...

Страница 3: ...lled horizontally Supply voltage 10 30V DC 0 10V output is powered by 24 V Power consumption 485 0 12W 4 20mA 0 5V 0 10V 1 2W Range Wind speed 0 40m S customizable 0 5m starting wind speed Wind direction 0 360 Accuracy Wind speed 0 5 2 FS Wind direction 3 Resolution Wind speed 0 01 m s Wind direction 1 Working environment 40 80 0 95 RH Wind resistance 75 m s Response time 1S IP grade 485 type IP65...

Страница 4: ...al blue Wind direction signal positive green Negative wind direction signal ANALOG OUTPUT 1 Interface description Wide voltage power supply input 10 30V DC power supply For 0 10V output equipment the power supply is 24 v 2 Example of connection mode sensor sensor Analog signal collector Analog signal collector 4 wire wiring diagram 3 wire wiring diagram ...

Страница 5: ...lists the driver names of several different 485 converters 3 Computing method 3 1 conversion calculation of current mode output signal The measuring range is 0 40m s 4 20mA output When the output signal is 12mA the current wind speed is calculated The span of wind speed range is 40m s which is expressed by 16mA current signal 40m S 16mA 2 5m s ma that is the current changes 1mA and the wind speed ...

Страница 6: ...e of initial structure 4 bytes Address code 1 byte Function code 1 byte Data area n bytes Error check 16 bit CRC code Time to end structure 4 bytes Address code the address of the transmitter which is unique in the same communication network factory default is 0x01 Function code instruction function indication issued by the host Data area the data area is the specific communication data pay attent...

Страница 7: ...or configuration address Content Operation Definition 0000 H 40001 decimal Instantaneous wind speed Read only Real time value of wind speed expanded by 100 times 0001 H 40002 decimal wind direction Read only Real time value of wind direction integer 0 due north 90 due east 0002 H 40003 decimal Maximum wind speed Read only Maximum wind speed after power on of equipment expanded by 100 times 0003H 4...

Страница 8: ...20 Ω terminal resistance at the same time 6 USB to 485 driver is not installed or damaged 7 The equipment is damaged Fault phenomenon no output or output error Possible reasons 1 The range corresponding error leads to PLC calculation error Please refer to the technical indicators in the first part for the range 2 Wrong wiring mode or wrong wiring sequence 3 The power supply voltage is wrong for 0 ...

Отзывы: