COND SGA-A & SGA-D Manuel_EN Rev3.doc
Page 6 on 20
Rev: 03/08/2006
3.
SWITCH SETTINGS
Switch Positions
e.g. The switches in Figure 3.1 are depicted as ALL ON.
3.1.
Output Settings - Switch 4
Use switch 4 to select the required
output and, if required, the low pass
filter and 5V Excitation. (See Tables 3.1
and 3.2)
Figure 3.1.
3.1.1.
Output Option
Output Option
Input Range
4-20mA
0 - 20mA
4-20mA
0 - 20mA
0 - 10V
0 - 5V
±10V
±5V
+ Full Scale
20mA
20mA
20mA
20mA
10V
5V
10V
5V
↑
↑
↑
↑
↑
↑
↑
↑
↑
0 4mA
0mA
12mA
10mA
5V
2.5V
0V
0V
↓
↓
↓
↓
↓
↓
↓
- Full Scale
n/a
n/a
4mA Note 1 0mA Note 1
0V
0V
-10V
-5V
Table 3.1.
N.B. Full scale output on the voltage ranges is achieved with a bi-polar (±) input
Note 1 Negative inputs can be accommodated on the current output ranges by setting the ‘Zero’ switch SW2
to +50% (Table 3.8) and setting SW1 to twice the required mV/V setting (Table 3.6).
3.1.2.
Switch 4
Analogue Output and Excitation Voltage Options - SW4
SW4 1
2
3
4
5
6
7
8
±10V 0
↓
0
↓
0
↓
X X 1
↑
=Filter in 1
↑
Filter out
1
↑
=10V Exc 0
↓
=5V Exc
±5V 0
↓
1
↑
0
↓
X X 1
↑
=Filter in 1
↑
Filter out
1
↑
=10V Exc 0
↓
=5V Exc
0-10V 0
↓
1
↑
1
↑
X X 1
↑
=Filter in 1
↑
Filter out
1
↑
=10V Exc 0
↓
=5V Exc
0-5V 1
↑
1
↑
1
↑
X X 1
↑
=Filter in 1
↑
Filter out
1
↑
=10V Exc 0
↓
=5V Exc
0-20mA
X X X 0
↓
0
↓
1
↑
=Filter in 1
↑
Filter out
1
↑
=10V Exc 0
↓
=5V Exc
4-20mA
X X X 1
↑
1
↑
1
↑
=Filter in 1
↑
Filter out
1
↑
=10V Exc 0
↓
=5V Exc