Appendix : Heller Specific Software Features
9.1
SW Alarm Config tab
Figure 23: Software Alarm Config tab
The SW (software) Alarm Config tab allows the user to set alarms in software to be
used by a DDE (Dynamic Data Exchange). The settings here create links for the
DDE to interpret using the standard protocol as listed in 9.2 below. For single gas
instruments such as the Rapidox 1100ZR only the parameters for channel 1 are
used. For the Rapidox 1100ZR3 (multichannel) all three channels can have
software programmed alarms defined.
1)
Channel 1 Alarms:
Select an option from Enabled or Disabled to activate
the alarms on this channel. The default is Disabled.
2)
CH1 Alarm 1 Set Point:
The current units (% or ppm) are displayed by the
text and can be changed using the “Units” drop down menu. Enter a set
point for the first software alarm here. For example, for an alarm set point
at 30% O
2
, enter “30” into the edit box. Note that alarm 1 is triggered when
RISING above this value.
3)
CH1 Alarm 2 Set Point:
Enter a set point for the second software alarm
here. For example, for an alarm set point at 0.01% O
2
, enter “0.01” into the
edit box. Note that alarm 2 is triggered when FALLING below this value.
If you are working with a Rapidox 1100ZR3 with three gas inlets then repeat this
process for the remaining channels 2 and 3. Otherwise these additional boxes
remain greyed out.