® Display Unit Requirements
SpecMetrix
® will provide a display unit with at least the minimal requirements installed which are
Windows 7 64-bit (SP 1), 16GB Memory, and SSD storage.
Installation and Site Requirements
Utilities
The
SpecMetrix
® In-line system requires a
dedicated
120VAC/240VAC AC 15-amp grounded mains
supply adjacent to the Display Unit and the In-line Control Unit. This means a direct connection back to
the fuse Unit.
The instrumentation should not be spurred off a thyristor-
controlled heater or motor due to the electrical interference
this would cause.
The Control Unit and Display Unit can be run on the same 15-
amp circuit. The Display Unit requires a standard
120VAC/240VAC receptacle; the Control Unit can either be
directly connected via conduit or use a standard
120VAC/240VAC single-phase receptacle.
Ruggedized Probe Assembly