9-Troubleshooting
43
9.10 Export Menu Item Not Accessible
The
Export
menu item under
Tools
will be greyed out and not accessible if:
1. No compact flash card is installed in the system when it starts up. The system should
be powered down, the card installed and the system powered up again. Be aware that
inserting a Compact Flash card with the system powered up can damage the card.
2. The card is not recognized by the system. Sensors & Software recommends the
Sandisk Extreme professional grade series Compact Flash cards. These cards are
widely available at consumer electronics stores.
3. The card is improperly formatted and is not recognized by the system. Try reformatting
the card and restarting the system. The flash card can only be formatted as FAT or
FAT16. Users running Windows 2000 or XP will have the additional options to format
as FAT32 and NTFS, both of which will not work with Conquest. If, after reformatting,
the card is still not recognized, a new card of the type recommended above should be
used.
4. The compact flash card does not have enough free memory space to accept all the
data in the Export folder. The system should be powered down, the card ejected, all
files removed from the card, the card re-installed, the system powered up, and Export
attempted again.
9.11 PCD Menu Item Not Accessible
The
PCD
menu item under
Tools
will be greyed out and not accessible if the Conquest sensor
head does not contain the PCD sensor.
9.12 Constant Prompt to Perform System Test
If the operator is constantly prompted to run the
System Test
resulting in the inability to collect
data, turn off the IEP (
Image Enhancement Processing
) in the
Tools
menu. If the IEP is
disabled, allow the system to warm up for 5 minutes before starting data collection.
9.13 Creating a Test Line for Data Quality
One of the best ways of detecting problems with the Conquest system is compare data with data
collected previously along the same line.
Soon after receiving the system and getting comfortable with its operation, collect a line of data at
a convenient, easily accessible location. The line does not have to be too long but 1-2 metres (3-
6 feet) is a good guide. This data line should be saved electronically and perhaps plotted out on
paper and dated. The test line could be collected say, every 6 months and, by reviewing the
previous data, system problems can be detected early. As well, if there is a suspected problem
with the system, this test line could be collected and compared with earlier tests.
Содержание Conquest Series
Страница 1: ... Copyright 2012 Sensors Software Inc Drawing Number 2006 00112 05 ...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 8: ......
Страница 12: ...Table of Contents Conquest iv ...
Страница 14: ...1 Overview Conquest 2 ...
Страница 20: ...3 Packing Up Conquest Conquest 8 ...
Страница 32: ...4 Principles of Operation Conquest 20 ...
Страница 42: ...5 Line Scans and Grid Scans Conquest 30 ...
Страница 44: ...6 Help Conquest 32 ...
Страница 58: ...10 Care and Maintenance Conquest 46 ...
Страница 69: ...Appendix C Instrument Compatibility Conquest C 2 ...
Страница 71: ...Appendix D Safety Around Explosive Devices Conquest D 2 ...
Страница 75: ...Conquest Appendix E Conquest Glossary E 4 ...