To enable Limits click the “Enable Limits” checkbox. 

 

Channel Select 

The “Channel Select” option selects the control channel to be used with the limits 

feature. The user can select between torque and speed. 

 

Output Mode 

The “Output Mode” option selects the polarity of the active / triggered state. 

 

Active High

 

Active / triggered state is high, inactive is low.

 

 

Active Low

 

Active / triggered state is low, inactive is high. 

 

Samples Required Before Trigger 

The “Samples Required Before Trigger” parameter sets the number of continuous 

samples that must exceed the high/low limits, before the limit output is set to the 

active / triggered state. 

 

Use the “Samples Required Before Trigger” slider bar to select a value, the slider 

is scaled from 0 to 1000 in increments of 10. 

 

Low / High Limits 

The “Low” and “High” limits set the trigger thresholds that control the limit output. 

Low,  High  or  both  limits  must  be  selected  to  use  the  Limits  feature.  The  High 

value must not be lower than the Low value and vice versa. 

 

Enable  the  High/Low  limit  by  clicking  the  High/Low  checkbox,  then  use  the 

High/Low slider bar to select the trigger level. 

Содержание RWT420

Страница 1: ...hould be read carefully and the safety instructions observed before installing or operating the equipment related to this manual Apollo Park Ironstone Lane Wroxton Banbury Oxon OX15 6AY Tel 44 0 1869...

Страница 2: ...ion of Sensor Technology Trademarks TorqSense is a registered trademark of Sensor Technology Ltd TorqView is a trademark of Sensor Technology Ltd LabVIEW is a trademark of National Instruments Corpora...

Страница 3: ...llecting data 16 Optional Accessories 17 Analog Lead 17 Digital Lead RWT420 440 series ONLY 17 Digital Adaptor Lead RWT420 440 series ONLY 17 Additional Related Products and Transducer Resources 17 Tr...

Страница 4: ...0 series 1 x TorqView program installation CD RWT Configuration Software Manuals Appendices in PDF format USB driver The following items may also be included 12pin Lumberg female and or 12pin Lumberg...

Страница 5: ...the transducer shaft such as single membrane couplings and that these are correctly selected for the transducer rating and speed An undersized coupling will not transmit the torque while the high iner...

Страница 6: ...n the clockwise sense along the transducer shaft axis produces a positive polarity output signal while torque applied in the anticlockwise sense produces a negative polarity output signal If an optica...

Страница 7: ...ther the 2 5m Analog or Digital Lead s 15 way D type connector with a supply voltage of 12 32VDC on Pin 1 and Ground on Pin 2 Note It is imperative that the voltage does not drop below 12V at any stag...

Страница 8: ...ence of green and red flashes will indicate the results of each system test The following table lists the tests and there test order for each test the transducer will output either a GREEN for PASS or...

Страница 9: ...ak Mode Peak Mode is active if configured to do so the analog output will output a peak value Please refer to the Peak Input section for further details Analog Scaling The Analog Scaling status messag...

Страница 10: ...nalog pin is short circuit in current output mode the analog output pin is open circuit Check the analog connector wiring for Analog CH0 2 Analog Fault CH1 Analog Speed In voltage output mode the anal...

Страница 11: ...is a control signal and has two main purposes It controls what data gets applied to the torque analog output and it can be used to reset the peaks The peak input is activated by grounding pin 8 of th...

Страница 12: ...RWT410 RWT430 ORT231 Torque Auto Assign1 Scale 0 FSD Speed Fast RPM Scale 0 Max Speed ORT232 RWT411 RWT412 RWT431 RWT432 ORT233 Torque Auto Assign1 Scale 0 FSD Power Watts Scale 0 FSD x Max Speed 9 5...

Страница 13: ...ctory Default Bipolar 10V 0V Zero Advanced Model OPTN F ORT24x RWT42x RWT44x User Selectable Voltage Current User Selectable Factory Default Bipolar 10V 0V Zero Fail Output The fail output signals whe...

Страница 14: ...as more advanced processes like auto reset or triggering after a delay The limits module is configured using Transducer Control please refer to the Transducer Control manual for further details The li...

Страница 15: ...f 8 consecutive torque readings Fixed Zero A fixed value can be input and saved Fixed Zero is only available in Transducer Control Unless the zero is saved any zero offset applied will be lost after a...

Страница 16: ...zeroing the transducer Resetting transducer to factory default To reset the transducer to the factory default settings i e the functions set by the factory on purchase follow the steps below 1 Turn o...

Страница 17: ...Sensor Technology 2016 Page 16 RWT3745V Rev 2 Examples of reading collecting data For diagrammatic purposes only the RWT410 420 model is shown...

Страница 18: ...of 1 amp to supply peak start up current AC Mains Adaptor Power Supply PSU The AC Mains Adaptor Power supply is required to provide 12 32VDC to the RWT series Transducer and Transducer Signal Breakout...

Страница 19: ...Out 12 Pin male on transducer Pin Description Pin 1 12 32VDC input G 2 GND M 3 Analog GND J 4 Analog CH1 Speed V or I E 5 Analog CH0 Torque V or I C 6 Reserved K 7 Zero Input low zero A 8 Peak Input l...

Страница 20: ...15 way D type connector Pwr Out 12 Pin female on transducer Pin Description Pin 1 12 32 VDC Input G 2 GND E 3 RS232 RXD CAN LO A 4 GND 5 RS232 TXD CAN HI K 6 GND 7 USB C 8 USB L 9 USB 0V D 10 USB Powe...

Страница 21: ...Sensor Technology 2016 Page 20 RWT3745V Rev 2 Transducer Connections Analog Digital Do NOT supply any voltage to these ports Closure to ground only...

Страница 22: ...S EN 61326 1 2006 Electrical equipment for measurement control and laboratory use Type of Product Rotary Torque Transducer Product Name s RWT410 RWT411 RWT413 RWT420 RWT421 RWT430 RWT431 RWT433 RWT440...

Страница 23: ......

Страница 24: ...put signal Pulse output direct from opto switch TTL 5V square wave output is independent of any analog or digital processing Processing Method Update rate for analog and digital outputs Mode 1 Slow Me...

Страница 25: ...100 Nm Shaft Code CF DA DF DB DC DG DD DE EB EC ED EE Shaft Size Diameter 6 12 20 mm Torsional Stiffness 0 23 1 28 1 3 1 32 1 6 1 7 1 8 1 9 4 1 6 4 8 1 9 2 KNm rad Mass moment of inertia LV 0 45 5 96...

Страница 26: ...ax 175 225 265 350 500 Nm Shaft Code FA FB FC FD FE Shaft Size Diameter 30 mm Torsional stiffness 32 9 35 6 37 2 37 9 39 8 kNm rad Mass moment of inertia 138 9 143 1 147 7 151 9 174 2 x 10 6 kg m2 Max...

Страница 27: ...00 13000 Nm Shaft Code GE GA GB GC GD HA HB HC HF HG Shaft Size Diameter 50 75 mm Torsional Stiffness TBC TBC 199 2 TBC 214 1 TBC TBC 914 4 945 5 TBC kNm rad Mass moment of inertia TBC TBC 1330 TBC 14...

Страница 28: ...output 1v FSD Fixed A In place of Option B Voltage output 10v FSD Fixed C In place of Option B Customer Specified Voltage Output Fixed U In place of Option B User to specify range scale when ordering...

Страница 29: ...For connecting RWT to PC via USB Option G or RS232 Also needs Digital Lead ACC4 to connect to RWT Digital Lead Adapter Length 1m 15 Way D type female to CANbus and Power Connectors ACC 8 For connecti...

Страница 30: ...t the system performance Frequency response of the IES method The measure of the TorqSense system s response at the output to a signal of varying frequency at its input The frequency response is typic...

Страница 31: ...Transducer Control 5 User Manual Revision 5 December 2016...

Страница 32: ...ansducer 5 Using Transducer Control 7 Menu Tree 7 Transducer Configuration 8 User Configuration 8 Analog Setup 8 Digital Setup 15 Zero Offset 21 Transducer Test 22 Output Monitor 22 Transducer Status...

Страница 33: ...can also be used as a readout providing a simple text display that shows torque speed and power along with peak values and zero control Compatible Models Advanced models from the ORT240 RWT320 RWT340...

Страница 34: ...Insert the Sensor Technology Installation CD into your CD ROM drive The following screen will be displayed 2 Click the Install button next to Transducer Control 5 3 Follow the onscreen instructions 4...

Страница 35: ...Windows button then Programs All apps Sensor Technology Transducer Control 5 When Transducer Control starts the Search for connected transducers dialog box will be displayed The purpose of the dialog...

Страница 36: ...ucers connection method is known it is recommended that it be selected if it is unknown select AUTO Once an option has been selected click on the Search button Transducer Control will then search for...

Страница 37: ...he menu structure below outlines the functionality of the different menu options File Exit Closes Transducer Control Transducer Configuration User Configuration Configures the operational parameters o...

Страница 38: ...to the output and select a data range to be applied to the Analog Voltage Scaling Data Select The following list details the available internal data variables that can be assigned to the output Speed...

Страница 39: ...peed Fast Speed Slow Current speed value The slow fast descriptor refers to the method used to capture speed Speed Fast Peak Speed Slow Peak Highest speed value captured since the last reboot or reset...

Страница 40: ...ted internal variable based on the FSD rating programmed into the transducer i e 0 to FSD User Defined The User Defined option allows the user to adjust the Range Low and Range High sliders to select...

Страница 41: ...w the range selection works Torque 100Nm FSD Standard Range 0 Nm 100 Nm 10v Scale 10V 10V 0V 100Nm Zero 100Nm Torque Voltage Torque 100Nm FSD User Defined Range 50 Nm 100 Nm 10v Scale 10V 10V 0V 100Nm...

Страница 42: ...els The data and range values selected in the Analog Data Select section are scaled to values configured here Users can either select from a list of common predefined scales or select a custom scale u...

Страница 43: ...Transducer Control 5 User Manual RWT3746IM Revision 5 December 2016 Page 13 Predefined Current Scales Option F Current Output...

Страница 44: ...maximum 10V Voltage Swing voltage will be reduced This feature is useful if the user has an analog capture device that does not support negative voltages unipolar only but requires the direction of t...

Страница 45: ...nditions where there may be erratic torque and speed readings The filters apply to the base data captured and will affect both the digital and analog outputs Enable or disable the filter for torque or...

Страница 46: ...counter exceeds the Samples Required Before Trigger parameter the limit output will be set to the active state level configured using the output polarity If the auto reset has not been configured the...

Страница 47: ...s Required Before Trigger The Samples Required Before Trigger parameter sets the number of continuous samples that must exceed the high low limits before the limit output is set to the active triggere...

Страница 48: ...l channel value no longer exceeds the low or high limit threshold Timeout The Timeout option is an extension of the Auto Reset Limit option and can only be enabled when Auto Reset Limit is enabled Whe...

Страница 49: ...o run at very high baud rates as some systems especially embedded devices may not be able to generate such high baud rates Long cable lengths may also hinder the use of higher baud rates To accommodat...

Страница 50: ...ransducer The Update Transducer button will upload the settings to the Transducer Transducer Control will validate the settings before transmission to ensure that there are no errors If an error is de...

Страница 51: ...the transducer would be reading 50 The zero offset can be temporary overridden by the digital zero commands or the Normal Average Zero Input When the transducer is power cycled the saved zero offset w...

Страница 52: ...transducer zeros the torque readings by recording the current torque as an offset the offset is then subtracted from all future torque readings AV Zero The AV Zero button works in the same way as the...

Страница 53: ...rating normally System Warning Transducer has detected an error which requires user intervention System Error Transducer has detected a system fault Status Code The Status Code is the value that has b...

Страница 54: ...a high load resistance or open load High Zero Offset A torque zero offset greater than 25 of FSD is in use Any offset value is allowed but the warning is asserted because it is potentially a very dan...

Страница 55: ...ailed temperature sensor error indicates that the transducers internal shaft temperature sensor has malfunctioned When the temperature sensor fails the transducer will no longer compensate the torque...

Страница 56: ...Exceeded Count Count of the number of times the torque level has gone over FSD Maximum Recorded Maximum recorded torque value over FSD Temperature Exceeded Count Count of the number of times the temp...

Страница 57: ...USER MANUAL TSE3168V Rev 7 Apollo Park Ironstone Lane Wroxton Oxon UK OX15 6AY Tel 44 0 1869 238400 Fax 44 0 1869 238401 Email info sensors co uk Web www sensors co uk...

Страница 58: ...minimum of 1GB of RAM a minimum screen resolution of 1024 x 768 pixels a Pentium 4 or equivalent processor and a minimum of 650 MB free disk space Installing TorqView Step 1 Insert the TorqView instal...

Страница 59: ...transducers connected only on USB ports COM1 etc will search for transducers connected only on COM1 etc ports If it is known which port the transducer is connected then select it from the list and cl...

Страница 60: ...out Screen to diagnose The Function Bar also shows the serial number of the currently active transducer If two transducers are connected they can be changed using the Setup Screen Click EXIT TORQVIEW...

Страница 61: ...eak Torque is displayed in two ways If the dial is in Center Zero mode orange needles will display the peak value for both the CC and CCW directions If the dial is in Normal mode one orange needle wil...

Страница 62: ...he same data less temperature as the Dials Screen but as digital bars The blue coloured bar displays the actual data whereas the orange coloured bar as an underlay of the blue coloured bar displays pe...

Страница 63: ...lluminated to indicate that the chart has started and no light illuminated to indicate that the chart has stopped With the chart is stopped the horizontal slider can be used to analyse the data The Cl...

Страница 64: ...on of the file Click RECORD on the TorqView Function Bar to start recording To view saved files double click file at saved destination The file will open in Notepad or a related text viewer Data will...

Страница 65: ...t your options and click next to continue The first screen is where you select what data you would like to record and any comments you would like to add Once you have finished click Next You will then...

Страница 66: ...on from the drop down fill in any additional information and click next The next screen will show you a summary of how the record has been set up If you are happy with details click the ARM button The...

Страница 67: ...ata will save to file and you will be asked if you would like to repeat the test If you select yes the record wizard will start again if you select no you will be given the option of running TorqView...

Страница 68: ...er Manual has been carried out and TorqView was purchased with the transducer then TorqView reactivation is not required However if a factory default reset has been carried out and TorqView was purcha...

Страница 69: ...as exceeded normal operating conditions Temperature compensation may no longer effective or calibration valid Overscale Torque has been applied greater than the specified FSD Critical overscale Torque...

Отзывы: