background image

GLT Application User Manual 

 

 

 

 

 

 

 

 

SENSIT Technologies, LLC   

 

16 

 

5

 

Troubleshooting 

Further sections to be added as necessary.

 

5.1

 

Communication Error 

A communication error, either shown by a dialog box or the 

application simply returning to the devices screen, indicates 

that either the tablet or the instrument has lost 

communication and the connection needs to be 

reestablished.  This could happen for a variety of reasons, 

but generally can be attributed to the system being used 

outside of the operating specifications.  Namely: 

1.

 

The tablet and instrument are more than 100ft away from each other. 

2.

 

There are a significant amount of line of sight (LOS) breaks. 

3.

 

The antenna(s) are being obstructed or covered (such as by the operators hand). 

If simply correcting an issue as listed above and trying to reconnect to the instrument through the application does not 

work, or a communication error is still being presented, close the application and restart. 

 

To close the application, select the task switching icon in the taskbar at the bottom of the screen.  The display should 

show all running application.  Select the ‘X’ in the upper right corner of the GLT application tile. 

 

To restart the application, select the start menu icon in the taskbar at the bottom of the screen.  Select the GLT icon 

from the start menu. 

 

If this does not resolve the issue, try to repair/re-initialize the Bluetooth connection (section 5.1.1 below).  If this still 

does not resolve the issue, contact Sensit for assistance. 

5.1.1

 

Repairing / Initializing Bluetooth connection 

In some cases, it may be necessary to re-establish the Bluetooth pairing between the instrument and tablet in order to 

resolve a communication issue.  To do so: 

NOTE:  Refer to section 4.1.2 

Pairing a New Instrument

 

 

for instructions on how to enable and disable the Bluetooth 

Support Service, which may need to be done in order to remove/pair devices.

 

1.

 

Un-pair and remove the device from the Windows Bluetooth manager. 

a.

 

Open the start menu by selecting the Windows icon in the lower left corner of the screen. 

b.

 

Select Bluetooth settings from the start menu icons in the middle of the screen. 

c.

 

On the right section of the menu, you should see a list of devices.  Select the instrument you are 

troubleshooting, and select remove device.  This will un-pair and remove the instrument from the device 

list. 

2.

 

If the instrument is on, power it down and restart it. 

a.

 

When the instrument gets to the “WAIT FOR TRAINER” screen, do not press any buttons. 

3.

 

On the tablet, wait for the instrument to reappear in the device list. 

a.

 

If it does not reappear, close out of the settings menu and re-open it by going back to step 1a. 

4.

 

Select the instrument and select the pair option. 

5.

 

Wait for the connection to establish.  The device should first list as “Connected”, followed by “Paired”. 

Содержание GASLESS LEAK TRAINER

Страница 1: ...M A N U A L 851 Transport Drive Valparaiso IN 46383 8432 Phone 888 473 6748 219 465 2700 Fax 219 465 2701 Email info gasleaksensors com Web www gasleaksensors com S U P P O R T S S E N S I T G O L D...

Страница 2: ...1 1 Multiple Instrument Support 6 4 1 2 Pairing a New Instrument 6 4 2 USER INTERFACE 8 4 3 USERS MENU 10 4 4 OQ TEST REPORTS 12 4 4 1 Performing an OQ Test 12 4 4 2 Generating a Report 13 4 5 SETTING...

Страница 3: ...se which allows for well organized data and easy retrieval and backup The Sensit instrument used with the GLT system is a fully functional unit with all of the same features as a standard one per the...

Страница 4: ...iners by both name and employee ID number Ability to assign instruments to specific users OQ operator qualification testing and reporting with the ability to customize questions 6 hour battery life wi...

Страница 5: ...up process a screen will be displayed alerting the user to connect the GLT software to the instrument Pressing any button on this screen will skip this process and boot the instrument into monitoring...

Страница 6: ...and other effects The instrument does not have the ability to detect and display gas in this mode only commands and simulated gas readings sent by the trainer Note The instrument must be restarted in...

Страница 7: ...powered off or the X button in the upper right of the screen is selected 4 1 2 Pairing a New Instrument If bought as a paired system tablet and instrument s together there should be no need to pair t...

Страница 8: ...lower left of the screen Select the GLT app icon from the middle of the screen 11 Resume the start up process as normal Refer to step 4 from section 4 1 Disabling Bluetooth Support Service 12 From the...

Страница 9: ...ing mode FLOW BLOCK sends a flow block command to the instrument BATT LOW sends a low battery command flag to the instrument A 300 second countdown will begin SET CAL DUE flags the instrument for cali...

Страница 10: ...MODE Slide bars allows the trainer to change concentrations of gas displayed on the instrument NOTE not available in monitoring mode Training module gives basic information on various displays and te...

Страница 11: ...d select save to add the user The new user will appear on the main display in the users menu Each instrument may be assigned to a user 4 To edit user information select the edit icon for the appropria...

Страница 12: ...igned to a user To do so select the assign user icon from the devices screen A box will appear to select a user Select the appropriate user followed by the OK button The user s name will appear under...

Страница 13: ...tion 4 3 1 Navigate to the user menu tab by selecting the icon on the left side of the application 2 To the right of each user s name and ID number will be a set of options including OQTest and Report...

Страница 14: ...ted by date and time in descending order 3 To generate and view a report select the HTML icon The report will be opened in either Microsoft Edge or whatever the default browser is set to 4 From there...

Страница 15: ...2 When the delete icon is selected a dialog box appears asking the user to confirm 3 When the add icon is selected a text entry box appears where the question text can be entered Select save when done...

Страница 16: ...ion button When selected it will open a File Explorer window in that directory The backup will be a data base file db with the following format GLTDD_Mon_YYYY_HH_MM_SS where D Mon Y represent the date...

Страница 17: ...menu If this does not resolve the issue try to repair re initialize the Bluetooth connection section 5 1 1 below If this still does not resolve the issue contact Sensit for assistance 5 1 1 Repairing...

Страница 18: ...ensit for further assistance 6 Software Update As of software revision 4 0 6 the GLT application may be updated through the Windows Store and does not need to be sent in to Sensit to be updated If you...

Страница 19: ...GLT Application User Manual SENSIT Technologies LLC 18 7 Notes...

Страница 20: ...t This warranty gives you specific legal rights and you may have other rights which vary from state to state Sensit Technologies LLC 851 Transport Drive Valparaiso Indiana 46383 USA Tel 219 465 2700 F...

Отзывы: