20
REF 360-0040-01 (F)
GILAIR-3 & GILAIR-5 AIR SAMPLING PUMPS
3.7 Retrieving “LAST” Run Data: Program Model
Note: Side Anti-Tamper Plate should be open and rotated.
1)
Make certain pump is off.
2)
Using a pointed instrument push MODE/HOLD Button one time.
3)
“LAST” is displayed.
NOTES
If CAL is displayed instead of LAST, press the MODE/HOLD button twice more to display the LAST run data.
If you turn on the pump the LAST run data will be erased and replaced by new run data. If the On/Off switch is set to On at this
point to begin a new sampling run, the “LAST” run data will replace the “Old” run data, which will be lost at that time, as the
last run data is cleared for the new sampling run data. However, you may retrieve the “Old” run data later by pressing the MODE/
HOLD button (with the On/Off switch in the Off position) until the “Old” data are displayed.
4)
Display then shows the program number as “Px,” where “x” can be the letter C or 1 through 6.
5)
The run time is then displayed,
“R”
, and the percentage of time the last sampling run varied more than
±
5% from set flow rate. Record the LAST run data at this time.
The display continues alternating among these three screens for 5 minutes, then shuts off.
1
2
3
4
5
PROG
GilAir/P
Program Timer
Gilian
®
MODE/HOLD
PROG
GilAir/P
Program Timer
Gilian
®
MODE/HOLD
LAST
PROG
GilAir/P
Program Timer
Gilian
®
MODE/HOLD
489
0
R
PROG
GilAir/P
Program Timer
Gilian
®
MODE/HOLD
P1
Battery
Fault
Flow Adj.
On
Off