TOTAL CHLORINE
REMOVE STRIP
Remove one eXact
®
Strip Micro CL (DPD-4) Total
Chlorine and set in a dry, convenient place. Replace
cap on bottle.
5
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip. The display
will automatically start to count up for 120 seconds.
*NOTE: Standard Method (4500-Cl G, procedure for total chlorine) requires
the reading to be made after 2 minutes from the time the KI is added. For
compliance testing, you must time the two minutes and then make your
measurement. NOTE: From testing in our lab, water samples above 70°F
(20°C), generally, reach a stabilized reading quicker than 2 minutes.
6
READ RESULTS
READ
result displayed as Total Chlorine. This result is
automatically stored in the TC2 menu. After testing is
complete, rinse the sample cell immediately and clean
with brush.
7
FILL, DIP, READ
If result is 6.0 or greater, repeat steps 5-7 above
using same sample and a fresh strip.
TIP