FREE CHLORINE
REMOVE STRIP
Remove one eXact
®
Strip Micro CL (DPD-1) Free
Chlorine and set in a dry, convenient place. Replace
cap on bottle
5
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip.
6
READ RESULTS
READ
result displayed as Free Chlorine. This
result is automatically stored in the CL1 menu.
After testing is complete, rinse the sample cell
immediately.
7
Important!
Do not empty
the CELL if you are also
testing for Combined Chlorine. You will need the sample.
FILL, DIP, READ
If result is 6.0 or greater, repeat steps 5-7 above
using same sample and a fresh strip.
TIP