7
ⓒ
Copyright SENKO CO., LTD.
LED
Zero: RED / Span: Orange
LEL
CO
H2S
O2
Max
Range
100% LEL
500 PPM
200 PPM
30% vol.
Gas for
Calibration
ZERO
SPAN
0% LEL
25% LEL
0 PPM
100 PPM
0 PPM
50 PPM
0 % vol.
20.9 % vol.
Flow rate
300 cc/min
300 cc
300 cc
300 cc
Time needed
for
calibration
ZERO: at once
SPAN: 120s
ZERO: at once
SPAN: 120s
ZERO: at once
SPAN: 120s
ZERO: 120s
SPAN: 120s
NOTICE
We recommend calibration once every six months, at least once a year. Sensor
’s output going down
as time goes by; measurement error will increase.
Calibration is crucial for the sensor performance and persons relying on this product for their safety.
WARNING
Do not Fresh Air Calibration unless you are certain you are in fresh,
uncontaminated air; otherwise, inaccurate readings can occur which can
false indicate that a hazardous atmosphere is safe. If you have any doubts as to the quality of the
surrounding air, do not use the Fresh Air Calibration feature. The calibration check is required to verify
span accuracy. Failure to follow this warning can result in serious personal injury or death.
button
Содержание SI-100C
Страница 1: ...SENKO CO LTD SI 100C Operating Manual...
Страница 9: ...9 Copyright SENKO CO LTD Certification...
Страница 10: ...10 Copyright SENKO CO LTD...
Страница 11: ...11 Copyright SENKO CO LTD...
Страница 12: ...12 Copyright SENKO CO LTD...
Страница 13: ...13 Copyright SENKO CO LTD...